Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Access Systems: PowerPort Slim Implantable Port, indicated for patient therapies requiring repeated access to the vascular system. Product Codes: 1716000, 1716070

Recall number
Z-1300-2014
Recalling firm
Bard Access Systems
Product
PowerPort Slim Implantable Port, indicated for patient therapies requiring repeated access to the vascular system. Product Codes: 1716000, 1716070
Status
Terminated
Initiated
2014-02-27
Posted
2014-03-27
Terminated
2014-09-05
Reason
Labeling discrepancy.The label states that the implantable port contains silicone suture plugs, however approximately 23% of the units contained in the affected lots have a PowerPort Slim titanium implantable port without silicone suture plugs.
Root cause
Packaging
Action
Notified by Urgent Customer Notification Letter date February 13, 2014, customers were asked to immediately examine their inventory and identify any product subject to this notification. In addition, if they have further distributed this product, please identify customers and notify them at once of this notification. Customers may include a copy of this letter in your notification to customers. Customers may choose to use the involved product with the understanding that the kit may contain a PowerPort Slim implantable port without suture plugs, or may return the applicable product for replacement or credit.
Quantity
777
Distribution
US Nationwide distribution including WA, KY, IL, and IA.
Lot, serial or model codes
Lot Numbers: REXA1302, REXA1350, REXB1160
510(k) numbers
not on file
PMA numbers
not on file
Product code
LJT
Firm FEI
3006260740
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register