FUJIFILM Healthcare Americas Corporation: FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Recall number
Z-1299-2026
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Status
Open, Classified
Initiated
2026-01-09
Posted
2026-02-05
Terminated
not on file
Reason
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Root cause
Process control
Action
On January 9, 2026 URGENT: MEDICAL DEVICE RECALL letters were emailed to customers. Actions to be taken: Please pay attention to the following points until the work is completed. " For safety reasons, please turn off the breaker of the distribution board to which the power cord of the unit is connected after turning off the power of the unit according to the operation manual. " Do not forget to turn off the breaker of the distribution board until the replacement of the contactor is completed. 1) Please post this notification on or near the device 2) Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 3) Please contact Technical Assistance Center (TAC) at 1-888-FUJIMED (1-888-385-4633) to open a service ticket. 4) Please report any known adverse events to FDA via MedWatch.
Quantity
16 units
Distribution
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Lot, serial or model codes
Model Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;
510(k) numbers
["K152294"]
PMA numbers
not on file
Product code
KPR
Firm FEI
1000513161
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register