FDA Medical Device Recalls (openFDA)
Stryker Communications: S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
- Recall number
- Z-1297-2021
- Recalling firm
- Stryker Communications
- Product
- S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
- Status
- Open, Classified
- Initiated
- 2021-01-29
- Posted
- not on file
- Terminated
- not on file
- Reason
- The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.
- Root cause
- Under Investigation by firm
- Action
- An Urgent Customer Notification Letter was mailed (certified) to the affected consignees on 01/29/2021. Customers will be notified by letter and a Stryker Representative will visit each account for product inspection and replacement. If customers notice any issues prior to inspection, customers should discontinue use and call Technical Support immediately at (866) 841-5663.
- Quantity
- 1,393 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.
- Lot, serial or model codes
- Serial Numbers: 2258046 2863490
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FQO
- Firm FEI
- 3002975342
- Firm city
- Flower Mound
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28