Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics, Locking Screw, T7, Catalog Number: 40-27622, Size: 2.7X22MM, Non Sterile,Titanium Alloy; Manufactured and Distributed by: Stryker Leibinger GmbH & Co. KG;Boetzinger Strasse 41; 0-79111 Freiburg, Germany

Recall number
Z-1297-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics, Locking Screw, T7, Catalog Number: 40-27622, Size: 2.7X22MM, Non Sterile,Titanium Alloy; Manufactured and Distributed by: Stryker Leibinger GmbH & Co. KG;Boetzinger Strasse 41; 0-79111 Freiburg, Germany
Status
Terminated
Initiated
2008-01-28
Posted
2008-03-25
Terminated
2008-06-04
Reason
Screws do not have locking threads
Root cause
Process control
Action
Recall notification letters (Urgent Product Recall) were sent on January 28, 2008 to all Stryker Branches/Agencies by Federal Express. All product distributed was requested to be located and returned, including from hospitals.
Quantity
100 units
Distribution
Nationwide-USA including states of MA, PA, GA, OH, and TX (hospital consignees) and Stryker Branches/Agencies in MA, NJ, PA, MD, VA, NC, GA, TN, KY, OH, MI, LK, OK, TX, and CA.
Lot, serial or model codes
Lot Numbers: H8W00F0B41; H9H00FB41, and H9R00F0B41
510(k) numbers
not on file
PMA numbers
not on file
Product code
HRS
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register