Skip to the content

FDA Medical Device Recalls (openFDA)

Getinge Group Logistics America, LLC: Operating Table System

Recall number
Z-1296-2021
Recalling firm
Getinge Group Logistics America, LLC
Product
Operating Table System
Status
Open, Classified
Initiated
2021-02-11
Posted
not on file
Terminated
not on file
Reason
There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.
Root cause
Device Design
Action
In February, 2021, Maquet GmbH/Getinge, will issue an Urgent Medical Device Correction notice to customers via FedEx 2 Day Delivery for the Magnus 1180 Operating Table due to a potential for hydraulic leakage to occur at the inclination function which may result in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. If this were to occur, it may lead to the patient having to be transferred to another OR table causing a delay in the procedure.
Quantity
4288 units
Distribution
Domestic: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. Foreign: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Luxembourg, Lybia, Macedonia, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom, Uzbekistan, Venezuela, Vietnam and Yemen.
Lot, serial or model codes
Part Number: 118001A0, 118001B0, 118001B1, 118001B2, 118001B3, 118001B4, 118001C0, 118001D0. UDI code: 04046768031500, 04046768031531, 04046768040700, 04046768088351, 04046768088061, 04046768124752, 04046768031555, 04046768031579. Serial Number: 1 3402, 1 1028, 1 1264, 1 401, 3 74, 1 81, 2 1159, 2 - 762.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FQO
Firm FEI
3025504317
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register