Abbott Laboratories, Inc: Abbott ARCHITECT cSystems ICT Module, REF 09D28-03. Product Usage: The ARCHITECT cSystems ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine.
- Recall number
- Z-1296-2019
- Recalling firm
- Abbott Laboratories, Inc
- Product
- Abbott ARCHITECT cSystems ICT Module, REF 09D28-03. Product Usage: The ARCHITECT cSystems ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine.
- Status
- Terminated
- Initiated
- 2018-08-28
- Posted
- not on file
- Terminated
- 2020-09-23
- Reason
- There is a potential to generate falsely elevated serum or plasma chloride results when using the product.
- Root cause
- Under Investigation by firm
- Action
- The recalling firm issued letters dated 8/28/2018 via overnight mail. The letter explained the issue, patient impact, and instructed the consignee to immediately discontinue use of the product and destroy any remaining inventory of the impacted lot number. The letter was to be reviewed with the Medical Director and laboratory protocol should be followed regarding the need to review previously-reported patient results. A Customer Reply Form was enclosed for the U.S. consignees to indicate the consignee understood and completed the necessary actions to be taken. For questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.
- Quantity
- 99 modules/kits
- Distribution
- Worldwide Distribution - US Nationwide in the states of AR, CA, CT, FL, IA, IN, LA, MA, MD, MS, NJ, NY, PA, SC, TX, VT, WA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Australia, China, Hong Kong, and New Zealand.
- Lot, serial or model codes
- Lot No. 180326, Exp. 12/26/2018; UDI (01)00380740019327(17)181226(10)180326(21)180326301 through (01)00380740019327(17)181226(10)180326(21)180326399
- 510(k) numbers
- ["K980367"]
- PMA numbers
- not on file
- Product code
- CGZ
- Firm FEI
- 1628664
- Firm city
- Irving
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28