Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories, Inc: Abbott ARCHITECT cSystems ICT Module, REF 09D28-03. Product Usage: The ARCHITECT cSystems ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine.

Recall number
Z-1296-2019
Recalling firm
Abbott Laboratories, Inc
Product
Abbott ARCHITECT cSystems ICT Module, REF 09D28-03. Product Usage: The ARCHITECT cSystems ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine.
Status
Terminated
Initiated
2018-08-28
Posted
not on file
Terminated
2020-09-23
Reason
There is a potential to generate falsely elevated serum or plasma chloride results when using the product.
Root cause
Under Investigation by firm
Action
The recalling firm issued letters dated 8/28/2018 via overnight mail. The letter explained the issue, patient impact, and instructed the consignee to immediately discontinue use of the product and destroy any remaining inventory of the impacted lot number. The letter was to be reviewed with the Medical Director and laboratory protocol should be followed regarding the need to review previously-reported patient results. A Customer Reply Form was enclosed for the U.S. consignees to indicate the consignee understood and completed the necessary actions to be taken. For questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.
Quantity
99 modules/kits
Distribution
Worldwide Distribution - US Nationwide in the states of AR, CA, CT, FL, IA, IN, LA, MA, MD, MS, NJ, NY, PA, SC, TX, VT, WA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Australia, China, Hong Kong, and New Zealand.
Lot, serial or model codes
Lot No. 180326, Exp. 12/26/2018;  UDI (01)00380740019327(17)181226(10)180326(21)180326301 through (01)00380740019327(17)181226(10)180326(21)180326399
510(k) numbers
["K980367"]
PMA numbers
not on file
Product code
CGZ
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register