Skip to the content

FDA Medical Device Recalls (openFDA)

Magellan Diagnostics, Inc.: LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.

Recall number
Z-1295-2018
Recalling firm
Magellan Diagnostics, Inc.
Product
LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
Status
Terminated
Initiated
2018-01-09
Posted
not on file
Terminated
2020-12-11
Reason
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Root cause
Device Design
Action
An Urgent Medical Device Correction letter, dated January 9, 2018, was sent to customers. The letter identified the affected device and provided a description of the issue. If two of the four AA batteries are inserted incorrectly, the batteries may overheat, causing a minor burn if the user opens the battery compartment and tries to remove the hot batteries. Customers are to follow the steps outlined in the letter, including placing the enclosed label on the upper back panel of the analyzer to remind users who are changing batteries of the importance of correct placement. Proper battery placement should be provided in annual training. Customers are to complete and return the acknowledgement card. Please direct any questions to 800-275-0102 or LeadCareSupport@magellandx.com.
Quantity
6,143
Distribution
Worldwide Distribution
Lot, serial or model codes
Part Numbers: 70-6760, 70-6529-R, 70-6760-P, and  70-6760-R. All LeadCare II and LeadCare Plus Blood Lead Analyzers with battery boards versions B and O.
510(k) numbers
["K052549"]
PMA numbers
not on file
Product code
DOF
Firm FEI
1218996
Firm city
North Billerica
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register