Skip to the content

FDA Medical Device Recalls (openFDA)

Transonic Systems Inc: Transonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis Monitor is intended for use by trained medical personnel to measure delivered blood flow, vascular access recirculation, vascular access blood flow and cardiac output on patients receiving hemodialysis treatment.

Recall number
Z-1295-2015
Recalling firm
Transonic Systems Inc
Product
Transonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis Monitor is intended for use by trained medical personnel to measure delivered blood flow, vascular access recirculation, vascular access blood flow and cardiac output on patients receiving hemodialysis treatment.
Status
Terminated
Initiated
2015-02-17
Posted
2015-03-18
Terminated
2016-05-31
Reason
Customer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user saw a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output).
Root cause
Process control
Action
Transonic sent a Medical Device Recall Letters and Response Forms dated February 17, 2015 to the users via e-mail and fax. The recall is being conducted to the user level. The letter identified the affected product, problem and actions to be taken. For questions call 607-257-5300 ext. 326.
Quantity
Domestic: 21 units, Foreign: 10 units
Distribution
Worldwide Distribution - USA Nationwide and the countries of: Australia, Canada, and South Korea
Lot, serial or model codes
Serial Numbers HD03A60004; HD03A81167; HD03A81225; HD03A81237; HD03A91338; HD03A91394; HD03A91406; HD03B01461; HD03B21782; HD03B21803; HD03A60020; HD03A71082; HD03A81207; HD03A91255; HD03A91316; HD03Bl1679; HD03B11684; HD03B11705; HD03B21812; HD03B31971; HD03A81240; HD03A71086; HD03B01459; HD03B01594; HD03B11667; HD03B42175; HD03B42160; HD03B42161; HD03A91258; HD03A70046; HD03B31901
510(k) numbers
["K960817"]
PMA numbers
not on file
Product code
MQS
Firm FEI
1319030
Firm city
Ithaca
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register