Skip to the content

FDA Medical Device Recalls (openFDA)

Seiler Instr. & Mfg. Co. Inc.: Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.

Recall number
Z-1294-2018
Recalling firm
Seiler Instr. & Mfg. Co. Inc.
Product
Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Status
Open, Classified
Initiated
2017-01-01
Posted
not on file
Terminated
not on file
Reason
The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.
Root cause
Use error
Action
The customer was notified via email on about January 4, 2017. Instructions included that the scope only tips if the user has it extended all the way out and is pulling on the handles of the scope. The end user must not pull on the handles of the scope when it is extended. They were also instructed to educate personnel on this matter. For further questions, please call (314) 968-2282.
Quantity
43 units, but 1 affected
Distribution
Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.
Lot, serial or model codes
Model Number 6 ENT
510(k) numbers
not on file
PMA numbers
not on file
Product code
EPT
Firm FEI
1927430
Firm city
Saint Louis
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register