FDA Medical Device Recalls (openFDA)

Murata Vios, Inc.: Vios Monitoring System Bedside Monitor Model BSM2050

Recall number
Z-1292-2025
Recalling firm
Murata Vios, Inc.
Product
Vios Monitoring System Bedside Monitor Model BSM2050
Status
Open, Classified
Initiated
2024-08-09
Posted
2025-03-06
Terminated
not on file
Reason
Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.
Root cause
Labeling design
Action
The firm sent a single page of instructions to customers on August 9, 2024 via email. Customers were informed that tablets should remain plugged into a wall outlet when the tablet is not in use. The tablet battery should not be allowed to fully deplete at any time during patient monitoring or when not in use.
Quantity
50
Distribution
US Nationwide distribution in the states of Florida, Louisiana, Maryland, Montana, New York, Ohio, and Texas.
Lot, serial or model codes
1.56.01 UDI: +B595BSM20500/$$7V1.56.01/16D20230630S 1.57.20 UDI: +B595BSM20500/$$7V1.57.20/16D20231129Y Software Versions 1.56.01 and 1.57.20
510(k) numbers
["K172586"]
PMA numbers
not on file
Product code
DRT
Firm FEI
3015328411
Firm city
Woodbury
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register