FDA Medical Device Recalls (openFDA)
Covidien LLC: Polysorb Braided Absorbable Suture 6-0 18" UNDYED SS-24 D/A, Item Code L2752K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in cardiovascular or neural tissue.
- Recall number
- Z-1292-2020
- Recalling firm
- Covidien LLC
- Product
- Polysorb Braided Absorbable Suture 6-0 18" UNDYED SS-24 D/A, Item Code L2752K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in cardiovascular or neural tissue.
- Status
- Terminated
- Initiated
- 2019-12-18
- Posted
- 2020-02-19
- Terminated
- 2022-08-16
- Reason
- There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Root cause
- Equipment maintenance
- Action
- On December 18, 2109, Medtronic issued Urgent Medical Device Recall and Return Form notices to customers via courier service. Customers were advised to take the following required actions: 1. Please immediately quarantine and discontinue use of the affected item code and lot listed within the notice. 2. Please return affected product as indicated within the notice. All unused products from the affected item codes and lots must be returned. 3. If you have distributed any of the Polysorb braided or Biosyn monofilament absorbable sutures listed on the recall notice, please promptly forward the information from the letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory.
- Quantity
- 139,938 units total
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Lot -D7H1092X Expiration Date - 2022-08-31
- 510(k) numbers
- ["K000037"]
- PMA numbers
- not on file
- Product code
- GAM
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28