Baxter Healthcare Corp.: The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Recall number
Z-1292-2015
Recalling firm
Baxter Healthcare Corp.
Product
The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Status
Terminated
Initiated
2013-10-16
Posted
2015-03-18
Terminated
2017-07-21
Reason
The occluder in affected devices 1) may have sharp edges that could cut the pneumatic tubing that connects the occluder blade to the manifold causing an unrecoverable system error, and/or 2) blade may separate from the back plate and cause Slow / No Flow alarm requiring therapy termination.
Root cause
Process design
Action
An Urgent Device Correction letter dated 10/16/13 was mailed to dialysis centers and medical facilities. A similar Urgent Device Correction letter dated 10/31/13 was mailed to dialysis home patients. The letters requested for customers to examine their devices to determine if the devices are being recalled. If the devices are on the attached list of recalled devices, the customer is to contact Baxter Technical Services at 1-800-553-6898 (7 days a week, 24 hours a day) to arrange for a swap cycler. The letter further requests that customers complete the attached Customer Reply Form and return it to Baxter by fax or e-mail.
Quantity
75 units (US)
Distribution
Worldwide Distribution-US (nationwide) including the states of AK, AR, AZ, CA, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, VT, WA, and WI, and the countries of Australia, Canada, and Mexico.
Lot, serial or model codes
Product Code: 5C8310;  Serial Numbers: 80415, 680416, 680418, 680420, 680421, 680422, 680424, 680426, 680427, 680428, 680429, 680430, 680438, 680439, 680440, 680441, 680444, 680445, 680448, 680449, 680452, 680453, 680454, 680455, 680456, 680457, 680459, 680460, 680461, 680462, 680463, 680465, 680467, 680467, 680469, 680470, 680472, 680473, 680475, 680476, 680477, 680478, 680480, 680481, 680482, 680463, 680484, 680485, 680486, 680489, 680490, 680491, 680494, 680495, 680496, 680498, 680499, 680500, 680501, 680502, 680503, 680504, 680505, 680506, 680507, 680508, 680509, 680510, 680516, 680517, 680519, 680520, 680521, 680522, 680524, 680525, 680526, 680527, 680527, 680528, 680529, 680530, 680532, 680533, 680534, 680536, 680537, 680538, 680539, 680541, 680542, 680543, 680547, 680552, 680553, 680544, 680555, 680561, 680562, 680566, 680567, 680585, 680586, 680587, 680588, 680589, 680590, 680591, 680593, 680594, 680595, 680596, 680597, 680598, 680599, 680600, 680601, 680602, 680603, 680604, 680605, 680606, 680607, 680608, 680609, 680611, 680612, 680613, 680614, 680615, 680616, 680617, 680618, 680619, 680620, 680621, 680622, 680623, 680624, 680625, 680626, 680627, 680628, 680629, 680630, 680631, 680632, 680633, 680634
510(k) numbers
["K102936"]
PMA numbers
not on file
Product code
FKX
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register