Skip to the content

FDA Medical Device Recalls (openFDA)

DeRoyal Industries Inc: DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement

Recall number
Z-1291-2025
Recalling firm
DeRoyal Industries Inc
Product
DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement
Status
Open, Classified
Initiated
2025-01-31
Posted
2025-03-05
Terminated
not on file
Reason
Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
Root cause
Under Investigation by firm
Action
On 01/31/2025, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via email to customers informing them that bulk roll, non-sterile suction connector tubing products and lots which could possibly be defective. The inner diameter of the suction tubing connector, is smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care. Customers are instructed to: 1) Identify affected product numbers in their inventory and place in quarantine to prevent further use. Destroy all product identified and complete the NOTICE OF RETURN FORM indicating all affected products found in their inventory. Return the completed form to DeRoyal via fax to 865-362-3716 or recalls@deroyal.com no later than March 7, 2025. Complete and return the Notice of Return Form even if you no longer have any invenorty. 2) If affected products have been further distributed to other facilities or departments within customer's institution, customers should be notified of this recall. by either directly forwarding the recall notification and notify DeRoyal that you have done so, or you may provide us with the customer listing and we will contact them. 3) Delay in patient care could lead to serious adverse events. A patient may aspirate, requiring immediate suction to prevent life-threatening injury. Inadequate suction not identified by the clinician could increase infection risk. For questions regarding credit for the recall, contact Jennifer Marsh at jmarsh@deroyal.com or by phone 865-362-6206 or email us at recalls@deroyal.com. If customers prefer to receive replacement product, contact DeRoyal Sales Representative or contact Angie Schubert, Assistant Product Director, at aschubert@deroyal.com.
Quantity
714 cases
Distribution
U.S.: IL, MN, NC, PA, and VA O.U.S.: N/A
Lot, serial or model codes
71-C5100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578738 Lot Number: CNWKD04-04 71-C6100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578851 Lot Numbers: CNWKD04-04 CNWKD04-05 CNWKD04-10 CNWKD09-08 71-C7100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578967 Lot Numbers: CNWKD04-06 CNWKD04-07
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAZ
Firm FEI
1043214
Firm city
Powell
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28