Skip to the content

FDA Medical Device Recalls (openFDA)

Aesculap Implant Systems LLC: DS Titanium Ligation Clip Appliers, used in laparoscopic surgery

Recall number
Z-1291-2018
Recalling firm
Aesculap Implant Systems LLC
Product
DS Titanium Ligation Clip Appliers, used in laparoscopic surgery
Status
Terminated
Initiated
2018-01-22
Posted
not on file
Terminated
2019-09-24
Reason
Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. Updated cleaning and sterilization instructions.
Root cause
Labeling design
Action
A letter and copy of the updated IFU was sent to affected customers the week of January 15, 2018. Product does not need to be returned. IFU is revised and put on the Aesculap Website for usage. https://www.aesculapusa.com/products/instructions-for-use.
Quantity
270
Distribution
US and Canada distribution
Lot, serial or model codes
DS Clip Appliers - Product Numbers PL808R DS SINGLE FIRE LAP.APPLIER L 12/310MM PL802R DS SINGLE FIRE LAP.APPLIER SM 5/310MM PL826R DS SINGLE FIRE LAP.APPLIER M 10/420MM PL817R DS SINGLE FIRE LAP.REM.FORC.ML 10/310MM PL828R DS SINGLE FIRE LAP.APPLIER L 12/420MM PL809R DS SINGLE FIRE LAP.APPLIER XL 12/310MM PL806R DS SINGLE FIRE LAP.APPLIER M 10/310MM PL818R DS SINGLE FIRE LAP.REM.FORC.L 12/310MM PL801R DS SINGLE FIRE LAP.APPLIER S 5/310MM PL816R DS SINGLE FIRE LAP.REM.FORC.M 10/310MM PL807R DS SINGLE FIRE LAP.APPLIER ML 10/310MM PL822R DS SINGLE FIRE LAP.APPLIER SM 5/420MM PL827R DS SINGLE FIRE LAP.APPLIER ML 10/420MM
510(k) numbers
["K080753","K102081"]
PMA numbers
not on file
Product code
FZP
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register