Skip to the content

FDA Medical Device Recalls (openFDA)

LUMENIS, LTD.: (1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.

Recall number
Z-1290-2025
Recalling firm
LUMENIS, LTD.
Product
(1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.
Status
Open, Classified
Initiated
2025-01-23
Posted
2025-03-05
Terminated
not on file
Reason
The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.
Root cause
Under Investigation by firm
Action
The recall was initiated via letter dated and issued on 1/23/2025 via FEDEX in the U.S. and via mail, courier, or hand delivery OUS. The letter explained the issue and the scenarios as to when the laser system would not power on. The consignee is instructed to immediately discontinue use and segregate the device, post this information in a visible location near the affected product to ensure the information is readily accessible to all handlers and users of the device, share the notice with any health care professional within your organization and with any organization where the affected product may have been transferred, and if the device has been further distributed, notify the customer to ensure this product removal is carried out to the end-user level. A Reply Verification Tracking Form was enclosed for completion and return. Once the firm receives the completed form, the consignee will be contacted within 2 business days and provided an RGA number for product return.
Quantity
3 devices
Distribution
Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.
Lot, serial or model codes
Software version 2.4.1.0. (1) REF GA-2009996 - Serial number 1791, UDI-DI 07290109145518; (2) REF GA-0006802 - Serial number 1803, UDI-DI 07290109145525; and (3) REF-GA0008700 - Serial number 1818, UDI-DI 07290109145464.
510(k) numbers
["K170121"]
PMA numbers
not on file
Product code
GEX
Firm FEI
3004135191
Firm city
Ilit
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register