Bard Peripheral Vascular Inc: BARD MARQUEE Disposable Core Biopsy Instruments XX.Xcm, XX.Xcm, XXg x XXcm REF XXXXXXX Note - for definition of the "X's refer to Tyvek Lid Label.pdf" uploaded in Associated Documents

Recall number
Z-1289-2021
Recalling firm
Bard Peripheral Vascular Inc
Product
BARD MARQUEE Disposable Core Biopsy Instruments XX.Xcm, XX.Xcm, XXg x XXcm REF XXXXXXX Note - for definition of the "X's refer to Tyvek Lid Label.pdf" uploaded in Associated Documents
Status
Terminated
Initiated
2021-01-14
Posted
not on file
Terminated
2025-01-28
Reason
Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.
Root cause
Process control
Action
On 02/11/2021, the firm sent an "UGENT MEDICAL DEVICE RECALL" Notification Letter via FedEx or email to customers/distributors informing them that the Recalling Firm has received reports that the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use and may results in (1) prolongation of the biopsy procedures as the user obtains a replacement device, (2) undershooting the lesion unsuspectedly, compromising the sample quality and potentially leading to a misdiagnosis (3) or if the penetration is overshot, the device may pierce adjacent tissue and/or organs causing injury. Customers are instructed to: 1. Check all inventory locations within their institution for the catalog/lot number combination listed in this recall notice and destroy all affected product remaining in their possession. 2. Share the recall notice with any users of the product, within their facilities, to ensure they are also aware of this recall. 3. If purchased this product from a distributor, contact the distributor for further instructions. 4. Complete and return the attached Customer Response Form to the Recalling Firm's contact noted on the form confirming acknowledgement of the recall notification, whether or not they have any affected product, so that Recalling Firm may acknowledge the receipt of the notification and process any applicable replacement orders. 5. Report any adverse health consequences experienced with the use of this product to the Recalling Firm. Actions to be taken by the Recalling Firm: 1. The firm will process replacement orders for all customers affected by the recall following receipt of the completed Customer Response Form. 2. Depending on the specific product code, there may be delays in the fulfillment of replacement orders. the firm understands that supply interruptions can impact our customers ability to provide the best care for their patients and takes t
Quantity
Total 131,876 kits (102,210 kits U.S. and 29,666 kits O.U.S.)
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY O.U.S.: Australia, Belgium, Canada, Europe, India, and Singapore
Lot, serial or model codes
Catalog and Product Description MQ1210 BARD MARQUEE Disposable Core Biopsy Instrument, 12g x 10cm,  MQ1213 BARD MARQUEE Disposable Core Biopsy Instrument, 12g x 13cm,  MQ1410 BARD MARQUEE Disposable Core Biopsy Instrument, 14g x 10cm,  MQ1413 BARD MARQUEE Disposable Core Biopsy Instrument, 14g x 13cm,  MQ1416 BARD MARQUEE Disposable Core Biopsy Instrument, 14g x 16cm,  MQ1610 BARD MARQUEE Disposable Core Biopsy Instrument, 16g x 10cm,  MQ1616 BARD MARQUEE Disposable Core Biopsy Instrument, 16g x 16cm,  MQ1620 BARD MARQUEE Disposable Core Biopsy Instrument, 16g x 20cm,  MQ1810 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 10cm,  MQ1816 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 16cm,  MQ1820 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 20cm,  MQ1825 BARD MARQUEE Disposable Core Biopsy Instrument, 18g x 25cm,  MQK1210 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 12g x 10cm,  MQK1213 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 12g x 13cm,  MQK1410 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 14g x 10cm,  MQK1413 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 14g x 13cm,  MQK1416 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 14g x 16cm,  MQK1610 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 16g x 10cm,  MQK1810 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 10cm,  MQK1816 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 16cm,  MQK1820 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 20cm,  MQK1825 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 18g x 25cm,  MQK2010 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 20g x 10cm,  MQK2016 BARD MARQUEE Disposable Core Biopsy Instrument Kit, 20g x 16cm  UDI Codes: (01)801741097119(17)230328(10)0001353340, (01)801741097119(17)230328(10)0001353342, (01)801741097119(17)230328(10)0001353401, (01)801741097119(17)230328(10)0001356645, (01)801741097119(17)230328(10)0001358065, (01)801741097119(17)230428(10)0001359983, (01)801741097119(17)230628(10)0001369253, (01)801741097126(17)230328(10)0001353402, (01)801741097126(17)230428(10)0001359984, (01)801741097126(17)230628(10)0001369256, (01)801741097133(17)230328(10)0001353403, (01)801741097133(17)230328(10)0001353404, (01)801741097133(17)230328(10)0001356410, (01)801741097133(17)230328(10)0001356409, (01)801741097133(17)230328(10)0001356886, (01)801741097133(17)230328(10)0001356888, (01)801741097133(17)230428(10)0001359150, (01)801741097133(17)230428(10)0001359152, (01)801741097133(17)230428(10)0001359149, (01)801741097133(17)230428(10)0001359151, (01)801741097133(17)230428(10)0001359988, (01)801741097133(17)230528(10)0001363603, (01)801741097133(17)230528(10)0001363604, (01)801741097133(17)230528(10)0001364078, (01)801741097133(17)230528(10)0001364076, (01)801741097133(17)230528(10)0001364077, (01)801741097133(17)230528(10)0001368474, (01)801741097133(17)230628(10)0001369045, (01)801741097133(17)230628(10)0001369046, (01)801741097140(17)230328(10)0001357081, (01)801741097140(17)230628(10)0001369273, (01)801741097157(17)230328(10)0001357079, (01)801741097164(17)230328(10)0001353246, (01)801741097164(17)230328(10)0001356414, (01)801741097164(17)230428(10)0001359989, (01)801741097164(17)230528(10)0001364079, (01)801741097164(17)230628(10)0001369264, (01)801741097171(17)230328(10)0001356999, (01)801741097171(17)230428(10)0001359990, (01)801741097188(17)230328(10)0001353723, (01)801741097195(17)230328(10)0001353492, (01)801741097195(17)230428(10)0001359991, (01)801741097195(17)230528(10)0001364091, (01)801741097195(17)230628(10)0001369267, (01)801741097201(17)230328(10)0001353722, (01)801741097218(17)230528(10)0001364092, (01)801741097218(17)230528(10)0001364093, (01)801741097218(17)230628(10)0001369270, (01)801741097225(17)230328(10)0001353721, (01)801741097225(17)230328(10)0001355807, (01)801741097225(17)230428(10)0001359994. (01)801741097225(17)230528(10)0001364090, (01)801741097263(17)230328(10)0001353400, (01)801741097263(17)230328(10)0001353399, (01)801741097263(17)230328(10)0001356646, (01)801741097263(17)230428(10)0001359178, (01)801741097263(17)230528(10)0001364069, (01)801741097263(17)230528(10)0001364070, (01)801741097263(17)230528(10)0001366667, (01)801741097263(17)230628(10)0001369252, (01)801741097263(17)230628(10)0001371860, (01)801741097270(17)230528(10)0001364072, (01)801741097270(17)230628(10)0001369259, (01)801741097287(17)230328(10)0001353727, (01)801741097287(17)230328(10)0001356411, (01)801741097287(17)230328(10)0001356412, (01)801741097287(17)230428(10)0001358408, (01)801741097287(17)230428(10)0001358409, (01)801741097287(17)230428(10)0001358414, (01)801741097287(17)230428(10)0001359999, (01)801741097287(17)230428(10)0001358413, (01)801741097287(17)230528(10)0001364084, (01)801741097287(17)230528(10)0001364085, (01)801741097287(17)230528(10)0001364082, (01)801741097287(17)230528(10)0001364080, (01)801741097287(17)230528(10)0001364081, (01)801741097287(17)230528(10)0001367194, (01)801741097287(17)230528(10)0001367195, (01)801741097287(17)230528(10)0001367196, (01)801741097287(17)230528(10)0001368556, (01)801741097287(17)230528(10)0001368555, (01)801741097287(17)230528(10)0001368760, (01)801741097287(17)230628(10)0001369292, (01)801741097294(17)230428(10)0001360001, (01)801741097300(17)230228(10)0001347094, (01)801741097300(17)230328(10)0001357065, (01)801741097317(17)230328(10)0001357859, (01)801741097317(17)230528(10)0001364088, (01)801741097317(17)230628(10)0001369278, (01)801741097348(17)230328(10)0001353725, (01)801741097348(17)230428(10)0001360028, (01)801741097348(17)230628(10)0001369289, (01)801741097355(17)230528(10)0001364089, (01)801741097355(17)230628(10)0001369284, (01)801741097362(17)230628(10)0001369290, (01)801741097379(17)230328(10)0001353724, (01)801741097379(17)230428(10)0001360029, (01)801741097386(17)230628(10)0001369272, (01)801741097393(17)230428(10)0001360030, (01)801741097393(17)230628(10)0001369276,  Lot Numbers: 0001353340, 0001353342, 0001353401, 0001356645, 0001358065, 0001359983, 0001369253, 0001353402, 0001359984, 0001369256, 0001353403, 0001353404, 0001356410, 0001356409, 0001356886, 0001356888, 0001359150, 0001359152, 0001359149, 0001359151, 0001359988, 0001363603, 0001363604, 0001364078, 0001364076, 0001364077, 0001368474, 0001369045, 0001369046, 0001357081, 0001369273, 0001357079, 0001353246, 0001356414, 0001359989, 0001364079, 0001369264, 0001356999, 0001359990, 0001353723, 0001353492, 0001359991, 0001364091, 0001369267, 0001353722, 0001364092, 0001364093, 0001369270, 0001353721, 0001355807, 0001359994, 0001364090, 0001353400, 0001353399, 0001356646, 0001359178, 0001364069, 0001364070, 0001366667, 0001369252, 0001371860, 0001364072, 0001369259, 0001353727, 0001356411, 0001356412, 0001358408, 0001358409, 0001358414, 0001359999, 0001358413, 0001364084, 0001364085, 0001364082, 0001364080, 0001364081, 0001367194, 0001367195, 0001367196, 0001368556, 0001368555, 0001368760, 0001369292, 0001360001, 0001347094, 0001357065, 0001357859, 0001364088, 0001369278, 0001353725, 0001360028, 0001369289, 0001364089, 0001369284, 0001369290, 0001353724, 0001360029, 0001369272, 0001360030, 0001369276
510(k) numbers
["K133948"]
PMA numbers
not on file
Product code
KNW
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register