FDA Medical Device Recalls (openFDA)
Ossur H / F: LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30
- Recall number
- Z-1288-2015
- Recalling firm
- Ossur H / F
- Product
- LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30
- Status
- Terminated
- Initiated
- 2015-03-04
- Posted
- 2015-03-17
- Terminated
- 2015-10-28
- Reason
- The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.
- Root cause
- Nonconforming Material/Component
- Action
- The firm's recall letter states the following instructions: Immediately examine inventory and quarantine product subject to the recall. In addition the firm asks if further distributed to identify the customers and notify them of the recall. The firm states that notification may be enhanced by providing a copy of the letter. The firm states that the recall notice needs to be passed along to all those who need to be aware. The firm also states that if customers are wearing a faulty product that they recommend them scheduling a visit to have the product replaced with a new LP Rotate that is currently available. The firm states that they can also arrange pick-up of defected products. For any questions the firm provides a customer service contact: Jamie Dutter, Senior Customer Service Representative.
- Quantity
- 2,194 (1,464 in U.S. and 730 internationally)
- Distribution
- Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,
- Lot, serial or model codes
- LP Rotate, sizes 25-27 cat.1-6. Product numbers LRP062, LRF062, LRPU62, LRFU62. All from manufacturing lot number HF141022
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- ISH
- Firm FEI
- 3003764610
- Firm city
- Reykjavik
- Firm state
- not on file
- Firm country
- Iceland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28