Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required); Model 801764. (Not distributed within the United States).

Recall number
Z-1287-06
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required); Model 801764. (Not distributed within the United States).
Status
Terminated
Initiated
2005-10-19
Posted
2006-07-27
Terminated
2008-06-19
Reason
While on battery power, the system may flash a low battery warning even though the battery is adequately charged.
Root cause
Other
Action
Customers were notified via letter dated October 19, 2005 that the Power Manager Software would be upgraded within the next six months. As of 7/7/06 all units within the U.S. have been completed and international units are still in the process of being upgraded.
Quantity
151
Distribution
Nationwide and worldwide, including Australia, Belgium, Canada, China, Dominican Republic, Egypt, Georgia, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand, Turkey and United Arab Emirates.
Lot, serial or model codes
All units using Power Manager software version 1.10, which includes serial numbers 6 through 156.
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register