Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories: CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

Recall number
Z-1286-2023
Recalling firm
Randox Laboratories
Product
CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826
Status
Open, Classified
Initiated
2023-02-01
Posted
2024-09-30
Terminated
not on file
Reason
The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range
Root cause
Under Investigation by firm
Action
Randox UK (manufacturer) extended Urgent Medical Device Removal letter to the US distributor via email on 2/1/23. The distributor contacted the customer via email on 2/6/23. Letter states reason for recall, health risk and action to take: " Discontinue use of and discard any of the above immediately. Provide Randox with photographic evidence of the destruction of the kits. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. If you have any questions or concerns, please contact Randox Technical Services.
Quantity
2 kits
Distribution
PA, OH, WV
Lot, serial or model codes
GTIN: 05055273201765 Batch: 588434
510(k) numbers
["K953212"]
PMA numbers
not on file
Product code
DCK
Firm FEI
3003059701
Firm city
Kearneysville
Firm state
WV
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register