Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: Field replacement back up batteries distributed on or after April 1, 2019 for Avance CS2, Avance CS2 Pro, Avance, Amingo, Aespire View anesthesia devices

Recall number
Z-1286-2022
Recalling firm
GE Healthcare, LLC
Product
Field replacement back up batteries distributed on or after April 1, 2019 for Avance CS2, Avance CS2 Pro, Avance, Amingo, Aespire View anesthesia devices
Status
Open, Classified
Initiated
2022-04-25
Posted
not on file
Terminated
not on file
Reason
Backup batteries can fail earlier than their estimated life.
Root cause
Under Investigation by firm
Action
An Urgent Medical Device Correction letter dated 04/25/2022 was sent to customers with affected devices. The letter mentions the users can continue to use the ventilator while connected to an AC mains power source that is supported by backup emergency power. The letter provides a series of recommendations for the consignee to follow. The letter also requests that the consignee complete the medical device notification acknowledgement response form and send it back to the firm.
Quantity
88,347 batteries
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, China, Columbia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Kazakhstan, Kenya, Korea, Kuwait, Lao PDR, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Nepal, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Lot, serial or model codes
Field Replacement Units (FRU) Battery kits - PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) or PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR), that may have been installed by the customer into any of the following anesthesia systems: o Avance 1009-9002-000 o Aespire View 1009-9212-000 o Avance CS2 and Avance CS2 Pro 1009-9050-000
510(k) numbers
["K172045"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register