Skip to the content

FDA Medical Device Recalls (openFDA)

Medos International Sarl: CERENOVUS CEREBASE DA Guide Sheath, Part Numbers: a) GS9080SD; b) GS9090SD; c) GS9095SD; Vascular guide-catheter, single-use

Recall number
Z-1284-2024
Recalling firm
Medos International Sarl
Product
CERENOVUS CEREBASE DA Guide Sheath, Part Numbers: a) GS9080SD; b) GS9090SD; c) GS9095SD; Vascular guide-catheter, single-use
Status
Open, Classified
Initiated
2024-02-02
Posted
2024-03-13
Terminated
not on file
Reason
Medos has received an increase in complaints for CEREBASE DA Guide Sheath with reports of fractures at the distal end, which may result in surgical procedural delay, vascular injury and in extreme rare occasions it may result in embolism.
Root cause
Under Investigation by firm
Action
Johnson&Johnson MedTech/DePuy Synthes issued an URGENT: MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 02/02/2024 via letter delivered by FedEx. The notice explained the issue and the risk and requested the affected inventory be quarantined and the notice be disseminated to all appropriate personnel. Further instructions state: "Complete the Business Reply Form (BRF) confirming receipt of this notice and scan and email signed form to OneMD-Field-Actions@its.jnj.com or fax it to (305) 265-6889, Attention Line: Cerebase FA2350411. Please return the BRF even if you do not have product subject to this recall. Follow instructions in the letter and immediately call 1-844-483-3882 to return any inventory of CEREBASE DA Guide Sheath devices subject to this recall. To receive credit reimbursement, customers must return product subject to this recall. Attention Line: Cerebase FA2350411"
Quantity
1343 units
Distribution
Worldwide distribution.
Lot, serial or model codes
a) GS9080SD: UPDATED 4/4/2024 UDI-DI 10886704082316, Lot numbers: 31121042, 31133636, 31140739, 31140751, 31146719, 31154809, 31160736, 31174927, 31178955, 31189206, 31194710, 31194733, 31208988, 31212653, 31212661; b) GS9090SD: UPDATED 4/4/2024 UDI-DI 10886704082293, Lot numbers: 31094249, 31103843, 31108260, 31116484, 31116485, 31121039, 31121040, 31121043, 31121044, 31133637, 31133638, 31140740, 31140952, 31146720, 31150968, 31150969, 31154810, 31154811, 31160737, 31160738, 31168610, 31168611, 31174928, 31174930, 31178956, 31178957, 31178958, 31189207, 31189208, 31194711, 31194712, 31194734, 31194735, 31204436, 31208989, 31208990, 31208991, 31212654, 31212655, 31212663, 31225738; c) GS9095SD: UDI-DI 10886704082323, Lot Numbers: 31208992, 31212656
510(k) numbers
["K192804"]
PMA numbers
not on file
Product code
QJP
Firm FEI
3008114965
Firm city
Le Locle
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register