Medtronic Vascular, Inc.: Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following designation: TFXXXXXXXXCP and TBXXXXXXXXCP. Medtronic Inc. Santa Rosa, CA 95403. The Talent Thoracic Stent Graft System is designed to treat thoracic aortic aneurysms using an endovascular approach. When placed within the target lesion, the stent graft provides an alternative conduit for blood flow within a patients vasculature by excluding the lesion from blood flow and pressure.

Recall number
Z-1284-2012
Recalling firm
Medtronic Vascular, Inc.
Product
Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following designation: TFXXXXXXXXCP and TBXXXXXXXXCP. Medtronic Inc. Santa Rosa, CA 95403. The Talent Thoracic Stent Graft System is designed to treat thoracic aortic aneurysms using an endovascular approach. When placed within the target lesion, the stent graft provides an alternative conduit for blood flow within a patients vasculature by excluding the lesion from blood flow and pressure.
Status
Terminated
Initiated
2012-02-21
Posted
2012-03-23
Terminated
2012-07-12
Reason
Complaints received related to the inability/difficulty to release the proximal (FreeFlow) bare stent using the tip capture mechanism.
Root cause
Process design
Action
Medtronic, Inc. sent an Urgent Medical Device Safety Information letter dated February 2012, to all affected customers. The letter identified the product, the problem, and the corrective action taken by Medtronic. Medtronic updated the Instructions for Use (IFU) manual. For questions regarding this recall call 707-566-1110.
Quantity
1523 units
Distribution
Nationwide Distribution
Lot, serial or model codes
TB2222C116CP, TF2424C116CP, TF2626C116CP, TF2828C116CP, TF3030C115CP, TF3232C115CP, TF3434C115CP, TF3636C114CP, TF3838C114CP, TF4040C114CP, TF4242C113CP, TF4444C113CP,  TF4646C112CP, TB2222C161CP, TF2424C161CP, TF2626C161CP, TF2828C161CP, TF3030C160CP, TF3232C160CP, TF3434C160CP, TF3636C159CP; TF3838C159CP, TF4040C159CP, TF4242C158CP, TF4444C158CP, TF4646C157CP, TB2222C201CP, TF2424C201CP, TF2626C201CP, TF2828C201CP, TF3030C200CP, TF3232C200CP, TF3434C200CP, TF3636C199CP, TF3838C199CP, TF4040C199CP, TF4242C198CP, TF4444C198CP, TF4646C197CP.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MIH
Firm FEI
3001452571
Firm city
Santa Rosa
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register