Skip to the content

FDA Medical Device Recalls (openFDA)

CardioTek BV: CardioTek EP-TRACER Software V2.x.

Recall number
Z-1283-2022
Recalling firm
CardioTek BV
Product
CardioTek EP-TRACER Software V2.x.
Status
Completed
Initiated
2022-04-25
Posted
not on file
Terminated
not on file
Reason
Device did not pass electrical safety testing for adequate insulation.
Root cause
Under Investigation by firm
Action
The firm issued a recall notification to consignees on 04/25/2022 via email. On 05/24/2022 the firm sent an updated field safety notice. The consignee can continue to use the device. The firm will contact the customer to schedule an appointment for the rework. The consignee is to pass on the letter to those that need to be made aware of the notice.
Quantity
146 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, NY, NC and the countries of South Korea, Mexico, Kazakhstan, Netherlands, Spain, Germany, Finland, Poland, Israel, Turkey, Chile, Ecuador, Austria, Croatia, Sweden, United Kingdom, Colombia, Italy, Portugal, Canada, Saudi Arabia, Belgium, Slovenia, Brazil, Hungary, Malaysia, Uzbekistan.
Lot, serial or model codes
EP-TRACER 38 UDI/DI: 04260441455418; Serial Numbers: 2019-1-01, 2019-1-02, 2019-1-03, 2019-1-04, 2019-1-05, 2019-1-06, 2019-2-01, 2019-2-02, 2019-2-03, 2019-2-04, 2019-2-05, 2019-3-01, 2019-3-02, 2019-3-03, 2019-3-04, 2019-3-05, 2019-3-06, 2019-3-07, 2019-3-08, 2019-3-09, 2019-3-10, 2019-4-01, 2019-4-02, 2019-4-03, 2019-4-04, 2019-4-05, 2019-4-06, 2019-4-07, 2019-4-08, 2019-4-09, 2019-4-10, 2019-5-01, 2019-5-02, 2019-5-03, 2019-5-04, 2019-5-05, 2020-1-01, 2020-1-02, 2020-1-03, 2020-1-04, 2020-1-05, 2020-1-06, 2020-1-07, 2020-1-08, 2020-1-09, 2020-1-10, 2020-1-11, 2020-1-12, 2020-1-13, 2020-1-14, 2020-1-15, 2020-2-01, 2020-2-02, 2020-2-03, 2020-2-04, 2020-2-05, 2020-2-06, 2020-2-07, 2020-2-08, 2020-2-09, 2020-2-10, 2020-2-11, 2020-2-13, 2020-2-14, 2020-2-15, 2020-4-01, 2020-4-02, 2020-4-03, 2020-4-04, 2021-1-01, 2021-1-02, 2021-1-03, 2021-1-04, 2021-1-05, 2021-1-06, 2021-1-07, 2021-1-08, 2021-2-01, 2021-2-02, 2021-2-03, 2021-2-04, 2021-2-06, 2021-2-07, 2021-2-08, 2021-2-09, 2021-2-10, 2021-2-11, 2021-2-12, 2021-2-13, 2021-2-14, 2021-2-15, 2021-3-01, 2021-3-02, 2021-3-03, 2021-3-04, 2021-3-05, 2021-3-06, 2021-3-07, 2021-3-08, 2021-3-09, 2021-3-10, 2021-3-11, 2021-3-12, 2021-3-13, 2021-3-14, 2021-3-15, 2021-5-01, 2021-5-02, 2021-5-03, 2021-5-04, 2021-5-05, 2021-5-06, 2022-1-01, 2022-1-03, 2022-1-04, 2022-1-05, 2022-1-06; EP-TRACER 102 UDI/DI: 04260441455234; Serial Numbers: 2019-1-11, 2019-2-11, 2019-2-12, 2019-2-13, 2019-2-14, 2019-2-15, 2019-2-16, 2019-4-21, 2019-4-22, 2019-4-23, 2019-4-24, 2019-4-25, 2019-4-26, 2019-4-27, 2020-1-26, 2020-1-27, 2020-1-28, 2020-1-29, 2020-1-30, 2020-2-26, 2020-2-27, 2020-2-28, 2020-2-29, 2020-2-30, 2021-1-13, 2021-1-14, 2021-2-27, 2022-1-26, 2022-1-27
510(k) numbers
["K183266"]
PMA numbers
not on file
Product code
DQK
Firm FEI
3008329188
Firm city
Maastricht-Airport
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register