Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc.: Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.

Recall number
Z-1282-2019
Recalling firm
Medtronic Inc.
Product
Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.
Status
Terminated
Initiated
2018-07-04
Posted
not on file
Terminated
2022-06-06
Reason
The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.
Root cause
Software design
Action
A Field Corrective Action was initiated outside of the U.S. by the recalling firm's global regions on 7/4/2018.
Quantity
Approximately 9,328 registered users
Distribution
Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.
Lot, serial or model codes
UDI - (01)00643169693142(10)
510(k) numbers
not on file
PMA numbers
["P160007"]
Product code
MDS
Firm FEI
3003166194
Firm city
Northridge
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register