FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc.: Atellica IM Erythropoietin (EPO) Assay (100 Test)
- Recall number
- Z-1280-2024
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Atellica IM Erythropoietin (EPO) Assay (100 Test)
- Status
- Open, Classified
- Initiated
- 2024-01-24
- Posted
- 2024-03-11
- Terminated
- not on file
- Reason
- Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
- Root cause
- Under Investigation by firm
- Action
- An Urgent Medical Device Correction notification letter dated 01/24/24 was sent to customers. Actions to be Taken by the Customer: Discontinue use of and discard the EPO kit lots listed in Table 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued EPO patient testing in your laboratory during this time, please contact your Siemens representative to discuss alternative Siemens Healthineers solutions. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
- Quantity
- 1,057 units (46 US, 1,011 OUS)
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, P.R. China, Pakistan, Poland, Portugal, Republic of South Korea, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, U.A.E., United Kingdom.
- Lot, serial or model codes
- Siemens Material Number: 10733006; UDI/DI: 00630414245775; Kit Lots: 53982039, 55439041.
- 510(k) numbers
- ["K183088"]
- PMA numbers
- not on file
- Product code
- GGT
- Firm FEI
- 2432235
- Firm city
- Tarrytown
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28