Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal Replacement Therapy (CRRT).

Recall number
Z-1280-2021
Recalling firm
Baxter Healthcare Corporation
Product
Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal Replacement Therapy (CRRT).
Status
Terminated
Initiated
2021-01-28
Posted
not on file
Terminated
2024-06-13
Reason
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Root cause
Process control
Action
An Urgent Medical Device Correction communication and reply form were sent to affected customers via U.S.P.S., first class mail on January 28, 2021.
Quantity
35175 kits
Distribution
Worldwide distribution - US Nationwide distribution including Puerto Rico and Guam and the countries of Albania, Algeria, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia Herzegovina, Bulgaria, Canada, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxemburg, Macedonia, Malta, Mauritius, Montenegro, Morocco, Netherlands, Nigeria, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sudan, Syria, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Yemen, Australia, New Zealand, China, Japan, Korea, Taiwan, Hong Kong, India, Thailand, Singapore, Malaysia, Vietnam, Grand Cayman, Dominican Republic, St. Kitts, Venezuela, Argentina, Chile, Barbados, Bermuda, Costa Rica, Cuba, Haiti, Guatemala, Colombia, Trinidad & Tobago, Mexico, and Brazil.
Lot, serial or model codes
all serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
KDI
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register