Skip to the content

FDA Medical Device Recalls (openFDA)

Christie Medical Holding Inc: VeinViewer (R) by Luminetx, Model Number VV1.0, Part No: P00399-L, Luminetx Corporation, Memphis, TN 38104 Product Usage: Non- invasive electronic visual aid device designed to project an image of a superficial subcutaneous vascular structures on the surface of the skin

Recall number
Z-1280-2012
Recalling firm
Christie Medical Holding Inc
Product
VeinViewer (R) by Luminetx, Model Number VV1.0, Part No: P00399-L, Luminetx Corporation, Memphis, TN 38104 Product Usage: Non- invasive electronic visual aid device designed to project an image of a superficial subcutaneous vascular structures on the surface of the skin
Status
Terminated
Initiated
2012-03-14
Posted
2012-03-22
Terminated
2012-10-23
Reason
Instrument may tip over if not moved correctly.
Root cause
Employee error
Action
Christie Medical Holdings, Inc. sent an Urgent Medical Device Correction by letter on March 14, 2012 to all affected customers. The letter identified the product, problem and actions to be taken. The letter includes "Correct Transport" and "Do Not Pull labels to be applied to the unit. Customers were instructed to immediately examine their inventory, quarantine all product subject to correction and apply the labels prior to use. A Service Bulletin is enclosed that should be kept and placed with the User Guide for future reference and a Response Form to be completed and return as soon as possible. For questions contact Technical Service at 901-721-0300.
Quantity
450 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Argentina, Australia, China, Brazil, Czech Republic, Canada, Germany, Denmark, Spain, Great Britain, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, South Korea, Kuwait, Mexico, Malaysia, Netherlands, Norway, Philippines, Russian Federation, Saudi Arabia, Singapore, Thailand, Turkey, Taiwan, and United Arab Emirates
Lot, serial or model codes
All codes distributed between 05/02/2006 08/31/2011
510(k) numbers
not on file
PMA numbers
not on file
Product code
KZA
Firm FEI
3008161930
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register