FDA Medical Device Recalls (openFDA)
Philips Medical Systems, Inc.: Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 This Philips MultiDiagnost Eleva is intended to be used as a multifunctional/universal system. General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Reproductive system, Respiratory system, Circulatory system, Various:Arthrograms, Myelograms, Facet joint injections, Discography, Sialography
- Recall number
- Z-1279-2016
- Recalling firm
- Philips Medical Systems, Inc.
- Product
- Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 This Philips MultiDiagnost Eleva is intended to be used as a multifunctional/universal system. General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Reproductive system, Respiratory system, Circulatory system, Various:Arthrograms, Myelograms, Facet joint injections, Discography, Sialography
- Status
- Terminated
- Initiated
- 2014-09-24
- Posted
- not on file
- Terminated
- 2019-03-26
- Reason
- Loss of key image functionality due to a bent pedal of the Footswitch.
- Root cause
- Device Design
- Action
- Field Safety Noticed was mailed on September 24, 2014 via Certified Return Receipt. This FCC consists of a Mechanical Hardware update of the Footswitch in order to prevent bent pedals and therefore prevents an intermittent or continuous inability of making live fluoro images or exposures. Customers with questions were instructed to contact their local Philips representative.
- Quantity
- 168
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA COLUMBIA COSTA RICA CROATIA CUBA CZECHOSLOVAKIA DENMARK DOMINICAN REPUB ECUADOR EGYPT ESTONIA FINLAND FRANCE GABON GERMANY GREECE HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KENYA KOREA SOUTH KUWAIT LATVIA LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVIA MONGOLIA MOROCCO NEPAL NETHERLANDS NETHERLANDS ANT NEW ZEALAND NORWAY OMAN PAKISTAN PALESTINE PANAMA Papua New Guinea PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUORTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERAT SAUDI ARABIA SERBIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SUDAN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TUNISIA TURKEY UKRAINE UNITED ARAB EMI United Arab Emirates UNITED KINGDOM URUGUAY USA UZBEKISTAN VENEZUELA VIETNAM YEMEN GUADELOUPE SYRIAN ARAB REPUBLIC
- Lot, serial or model codes
- Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2; 459800415531 Wireless Footswitch 3P; 459800415561 Wireless footswitch set 3P.
- 510(k) numbers
- ["K050151"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28