FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS Chemistry Products --- ALT Slides --- [REF] 165 5281 --- [Mfg Symbol] Ortho-Clinical Diagnostics, Inc., Rochester, NY USA. The recalled device is the calibration disk contained within the VITROS Calibrator Kit 3. The is can be identified with the label "VITROS Chemist y Systems --- Universal Calibration Diskette --- [REF] 871 6607 --- Ortho-Clinical Diagnostics" For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 3 is used to calibrate VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System for the quantitative measurement of AcP, ALT, ALKP, AMYL, AST, CK, GGT, LDH, and LIPA.
- Recall number
- Z-1277-2012
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry Products --- ALT Slides --- [REF] 165 5281 --- [Mfg Symbol] Ortho-Clinical Diagnostics, Inc., Rochester, NY USA. The recalled device is the calibration disk contained within the VITROS Calibrator Kit 3. The is can be identified with the label "VITROS Chemist y Systems --- Universal Calibration Diskette --- [REF] 871 6607 --- Ortho-Clinical Diagnostics" For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 3 is used to calibrate VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System for the quantitative measurement of AcP, ALT, ALKP, AMYL, AST, CK, GGT, LDH, and LIPA.
- Status
- Terminated
- Initiated
- 2012-02-02
- Posted
- 2012-03-22
- Terminated
- not on file
- Reason
- Ortho Clinical Diagnostics received complaints of lower than expected results (< 6 U/L) for samples that were known to be outside of the measuring (reportable) range (> 1000 U/L) when using Calibration Diskette/Assay Data Diskette Data Release Version (DRV) 5653 through 5666 and VITROS Chemistry Products ALT Slides, Generation (GEN) 26 (Product Code 1655281).
- Root cause
- Software design
- Action
- Ortho Clinical Diagnostic sent a Important Product Correction letter dated February 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed customers to: Install Calibration Diskette/Assay Data Diskette DRV 5667 on your VITROS System. Following a successful recalibration, you may resume using VITROS ALT Slides, GEN 26 remaining in your inventory. To please complete and return the Confirmation of Receipt form upon receipt of this notification. Please return this form no later than February 17, 2012. Post this notification by each VITROS System in your facility that utilizes the VITROS ALT Slides, or with the VITROS user documentation. Forward this notification if you have distributed this product outside of your facility. We apologize for the inconvenience this may have caused your facility. If you have any additional questions, please contact Customer Technical Services at 1-800-421-3311.
- Quantity
- 4306
- Distribution
- Worldwide Distribution - USA (nationwide)
- Lot, serial or model codes
- Calibration Diskette (Product Code 8716607) and the Assay Data Diskette (Product Code 6801876), Data Release Version (DRV) 5653 through 5666 used in conjuction with of VITROS¿ Chemistry Products ALT Slides, Generation (GEN) 26 (Product Code 1655281).
- 510(k) numbers
- ["K001679","K922072"]
- PMA numbers
- not on file
- Product code
- JIX
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28