Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190

Recall number
Z-1275-2022
Recalling firm
Olympus Corporation of the Americas
Product
EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190
Status
Open, Classified
Initiated
2022-04-22
Posted
not on file
Terminated
not on file
Reason
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications
Root cause
Nonconforming Material/Component
Action
Olympus issued Urgent Medical Device Correction letter via email on 4/22/22. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide for a service loaner in an effort to not disrupt patient care. Please do not hesitate to contact @ 647-999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Quantity
8 units
Distribution
US Nationwide distribution in the states of AL, FL, NC, NY.
Lot, serial or model codes
UDI-DI: 04953170305276 Serial: 2269931, 2611873, 2613235, 2634251, 2634326, 2860721, 2961773 and 2962545
510(k) numbers
not on file
PMA numbers
not on file
Product code
FDS
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register