FDA Medical Device Recalls (openFDA)
Synthes, Inc.: Drill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, including osteopenic bone in the ulna.
- Recall number
- Z-1274-2014
- Recalling firm
- Synthes, Inc.
- Product
- Drill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, including osteopenic bone in the ulna.
- Status
- Terminated
- Initiated
- 2014-01-07
- Posted
- 2014-03-25
- Terminated
- 2015-09-10
- Reason
- The Drill Template and the Saw Guide for the Ulna Osteotomy System may exhibit jamming, bending, or breaking of the screw connecting the Saw Guide to the Drill Template or jamming, bending, or breaking of the K-wire during fixation of the Drill Template.
- Root cause
- Component design/selection
- Action
- A recall notification letter, dated January 7, 2013, was sent to sales consultants and Users.
- Quantity
- 343
- Distribution
- Nationwide Distribution-including the states of AR, AK, VA, IN, PA, OH, CT, TX, MI, WI, NC, CO, IL, and MA.
- Lot, serial or model codes
- Part Nos. 03.111.900 through 03.111.905, with Lot Nos. 779372, 2638601, 7769386, 2638603, 7836420, 7772531, 2638604, 7836422, 7772538, 2638605, 7836423, 7915554, 7772543, 7772542, 7872775, 8400287, 7772554, 7772555, and 8229040.
- 510(k) numbers
- ["K113364"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 3005180112
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28