Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: REF: PIG1260K, 6Fr, X10, Safe-t-Centesis Catheter drainage Kit, Sterile EO

Recall number
Z-1273-2022
Recalling firm
Bard Peripheral Vascular Inc
Product
REF: PIG1260K, 6Fr, X10, Safe-t-Centesis Catheter drainage Kit, Sterile EO,
Status
Open, Classified
Initiated
2022-04-22
Posted
not on file
Terminated
not on file
Reason
There is the potential that a 6 Fr Catheter Drainage kit contains an 8 Fr needle instead of a 6 Fr needle. They can be used in conjunction with the following drainage methods: Universal extension set and drainage bag, vacuum bottle, and wall suction.
Root cause
Under Investigation by firm
Action
On March 18, 2022, Bard Peripheral Vascular issued a "Urgent Medical Device Correction" notification to affected consignees via E-Mail or Fed Ex. The notification ask consignees to take the following actions: 1. Please check all inventory locations within your institution for the product listed in Table A and immediately destroy all affected product remaining in your possession. The product should have the sterile barrier broken and destroyed per your local facility s destruction process. 2. Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. 3. If you purchased this product from a distributor, contact your distributor for further instructions and credit resolution. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement or credit. 5. Indicate on the response form the quantity of affected product identified at your facility and confirm that this inventory was destroyed. 6. Please contact your BD representative to assist in this process or you may also contact the number for further assistance. 7. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch; Phone: 1-800-FDA-1088 (1-800-332-1088); Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787
Quantity
1460 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of MS, NC, FL, KS, GA, OH, AZ, NV, OH, IL, MI, MO,IA, CA, PA, AL,WA, TN, IN, NY and the countries of AU, BE, NZ.
Lot, serial or model codes
Lot # 0001405396; UDI/DI: (01)10885403076114
510(k) numbers
not on file
PMA numbers
not on file
Product code
PXI
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register