Philips Healthcare (Suzhou) Co., Ltd.: Philips CT systems are advanced continuous-rotation computed tomography systems suitable for a wide range of computed tomographic (CT) applications - Product Usage: intended for use in the head and whole body.
- Recall number
- Z-1273-2021
- Recalling firm
- Philips Healthcare (Suzhou) Co., Ltd.
- Product
- Philips CT systems are advanced continuous-rotation computed tomography systems suitable for a wide range of computed tomographic (CT) applications - Product Usage: intended for use in the head and whole body.
- Status
- Terminated
- Initiated
- 2021-02-24
- Posted
- not on file
- Terminated
- 2023-02-17
- Reason
- Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.
- Root cause
- Process control
- Action
- On 2/18/2021, Philips issued an Urgent Medical Device Correction because the firm has detected an issue with some actuators used in the couch of Incisive CT and Access CT systems, which could result in the couch unexpectedly moving slowly downward to its limit position.
- Quantity
- 109 systems
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of KY, FL, St. Croix, and the countries of Australia, Bolivia, China, Germany, Hungary, Italy, Japan, Republic of Korea, Latvia, Palestine, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Spain, Sweden, Thailand, Uzbekistan.
- Lot, serial or model codes
- Unique Device Identifier (01)00884838059849(21)304034 (01)00884838059849(21)325078 (01)00884838085015(21)500218 (01)00884838085015(21)500219 (01)00884838085015(21)500221 (01)00884838085015(21)500222 (01)00884838085015(21)500223 (01)00884838085015(21)500224 (01)00884838085015(21)500225 (01)00884838085015(21)500226 (01)00884838085015(21)500227 (01)00884838085015(21)500229 (01)00884838085015(21)500231 (01)00884838085015(21)500233 (01)00884838085015(21)500234 (01)00884838085015(21)500237 (01)00884838085015(21)500238 (01)00884838085015(21)500239 (01)00884838085015(21)500241 (01)00884838085015(21)500242 (01)00884838085015(21)500245 (01)00884838085015(21)500247 (01)00884838085015(21)500248 (01)00884838085015(21)500249 (01)00884838085015(21)500251 (01)00884838085015(21)500252 (01)00884838085015(21)500254 (01)00884838085015(21)500255 (01)00884838085015(21)500256 (01)00884838085015(21)500257 (01)00884838085015(21)500258 (01)00884838085015(21)500261 (01)00884838085015(21)500264 (01)00884838085015(21)500274 (01)00884838085015(21)500283 (01)00884838085015(21)500296 (01)00884838085015(21)530038 (01)00884838085015(21)530067 (01)00884838085015(21)530071 (01)00884838085015(21)530098 (01)00884838085015(21)530099 (01)00884838085015(21)530101 (01)00884838085015(21)530103 (01)00884838085015(21)530104 (01)00884838085015(21)530106 (01)00884838085015(21)530107 (01)00884838085015(21)530108 (01)00884838085015(21)530109 (01)00884838085015(21)530111 (01)00884838085015(21)530112 (01)00884838085015(21)530113 (01)00884838085015(21)530114 (01)00884838085015(21)530115 (01)00884838085015(21)530116 (01)00884838085015(21)530117 (01)00884838085015(21)530118 (01)00884838085015(21)530119 (01)00884838085015(21)530120 (01)00884838085015(21)530121 (01)00884838085015(21)530124 (01)00884838085015(21)530126 (01)00884838085015(21)530127 (01)00884838085015(21)530128 (01)00884838085015(21)530129 (01)00884838085015(21)530130 (01)00884838085015(21)530132 (01)00884838085015(21)530133 (01)00884838085015(21)530134 (01)00884838085015(21)530135 (01)00884838085015(21)530136 (01)00884838085015(21)530137 (01)00884838085015(21)530138 (01)00884838085015(21)530141 (01)00884838085015(21)530148 (01)00884838085015(21)530150 (01)00884838085015(21)530152 (01)00884838085015(21)530156 (01)00884838085015(21)530157 (01)00884838085015(21)530159 (01)00884838085015(21)530160 (01)00884838085015(21)530163 (01)00884838085015(21)530165 (01)00884838085015(21)530166 (01)00884838085015(21)530167 (01)00884838085015(21)530169 (01)00884838085015(21)530170 (01)00884838085015(21)530171 (01)00884838085015(21)530172 (01)00884838085015(21)530173 (01)00884838085015(21)530174 (01)00884838085015(21)530175 (01)00884838085015(21)530176 (01)00884838085015(21)530177 (01)00884838085015(21)530178 (01)00884838085015(21)530180 (01)00884838085015(21)530181 (01)00884838085015(21)530182 (01)00884838085015(21)530183 (01)00884838085015(21)530186 (01)00884838085015(21)530187 (01)00884838085015(21)530188 (01)00884838085015(21)530189 (01)00884838085015(21)530190 (01)00884838085015(21)530191 (01)00884838085015(21)530192 (01)00884838085015(21)530194 (01)00884838085015(21)530197 (01)00884838085015(21)530198 (01)00884838085015(21)500230
- 510(k) numbers
- ["K180015"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 3009529630
- Firm city
- Suzhou
- Firm state
- not on file
- Firm country
- China
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28