Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.

Recall number
Z-1273-2020
Recalling firm
Smith & Nephew, Inc.
Product
TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.
Status
Terminated
Initiated
2019-12-30
Posted
not on file
Terminated
2021-05-11
Reason
Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.
Root cause
Package design/selection
Action
Smith Nephew "Urgent Medical Device Recall Notice R-2019-22" letter dated December 30, 2019. Customers were advised to: 1.Please inspect your inventory and locate any devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Quantity
82 distributed throughout US:AZ, ID, MO, NC, ND, NE, OK, SC, TN. 243 distributed OUS: AE, AU, CA, CH, ES, FI, JP, PR, SG. Total: 325 units
Distribution
Worldwide distribution US Nationwide distribution in the states of AZ, ID, MO, NC, ND, NE, OK, SC, TN, and countries of AE, AU, CA, CH, ES, FI, JP, PR, SG.
Lot, serial or model codes
Batch numbers: 2010547, 2025137, 2036695, 50763859, & 50773071. Part Number 72200796. UDI 10600147.
510(k) numbers
["K152566","K972326"]
PMA numbers
not on file
Product code
MBI
Firm FEI
1219602
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register