Skip to the content

FDA Medical Device Recalls (openFDA)

Vapotherm: Low Flow Disposable Patient Circuit for the Precision Flow System, Product Code PF-DPC-LOW

Recall number
Z-1272-2023
Recalling firm
Vapotherm
Product
Low Flow Disposable Patient Circuit for the Precision Flow System, Product Code PF-DPC-LOW
Status
Open, Classified
Initiated
2022-12-16
Posted
2023-03-17
Terminated
not on file
Reason
There have been reports of excessive condensation associated with certain lots of the Low Flow Disposable Patient Circuits (DPCs) used with Precision Flow System which may lead to excessive rainout.
Root cause
Nonconforming Material/Component
Action
A Technical Bulletin dated 10/27/22 was distributed to customers. Vapotherm has learned of certain Low Flow Disposable Patient Circuit (DPC) lots exhibiting excessive condensation. Please find the affected DPC lots in the attached appendix. You have been identified as a customer that received some or all of these lots. This condensation can be mitigated by ensuring that the delivery tube is fully primed and warm before placing the therapy on the patient, per existing instructions for use. Always check that the delivery tube is warm across the entire length before placing the therapy on the patient. If you experience delivery tubes that are not circulating water or are cold, do not use. Please contact Vapotherm Technical Support to report the issue. For every incident reported, we will replace and ship out a new DPC as soon as possible at no cost to you. Thank you for your continued support of Vapotherm and providing your patients with optimal care. If you have any questions, please contact Vapotherm Technical Support: For US customers, ts@vtherm.com or 855 557 8276 For UK customers, ukinfo@vtherm.com or +44 0345 305 8098 For all other customers, please reach out to your Vapotherm distributor. An Urgent Field Safety Notice dated 12/16/22 was sent to customers in the UK. Vapotherm requests that you return any unused products of the lots detailed below. Unused products from the affected lots should be returned as described in the Required Actions section within the FSN. This product removal is limited to the product codes / lot numbers listed below. No other product codes or lot numbers are affected. Required Actions to be taken by the VAPOTHERM: 1.For any unused product of the affected lots, replace affected DPC lots. 2.Email Customer Acknowledgment Form to impacted customers. 3.Ensure completion of Customer Acknowledgement Form. Required Actions to be taken by the END USER/ HEALTHCARE ORGANIZATION: 1.Inspect your inventory, locate any unused units of the af
Quantity
5869 total
Distribution
Worldwide distribution - US Nationwide and the country of UK.
Lot, serial or model codes
UDI-DI: 00841737104971, Lot Numbers: LFVMA8503, LFVMA8504, LFVMA8567; UDI-DI: 00841737104978, Lot Number: LFVLE8568 Additional Lot Numbers: LFTL41015 LFVLC8399 LFVLE8495 LFVMA8502 LFVMA8503 LFVMA8504 LFVMA8567 LFTL41261 LNTL41544 LFVLE8588 LFVMA8507 LFVLE8568 LFVLE8494 LFTL41263 LFTL41262
510(k) numbers
["K111640"]
PMA numbers
not on file
Product code
BTT
Firm FEI
3012971441
Firm city
Exeter
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register