SIE AG, Surgical Instrument Engineering: The FEMTO LDV(TM)Z8 Femtosecond Surgical Laser The FEMTO LDV(TM)Z8 Femtosecond Surgical Laser is an ophthalmic surgical laser intended for use in the creation of corneal incisions indicated for use in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty, penetrating keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface. In addition, the FEMTO LDV(TM)Z8 Surgical Laser is intended for use in the creation of capsulotomy, phacofragmentation and the creation of single plane, multi-plane, arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.

Recall number
Z-1272-2018
Recalling firm
SIE AG, Surgical Instrument Engineering
Product
The FEMTO LDV(TM)Z8 Femtosecond Surgical Laser The FEMTO LDV(TM)Z8 Femtosecond Surgical Laser is an ophthalmic surgical laser intended for use in the creation of corneal incisions indicated for use in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty, penetrating keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface. In addition, the FEMTO LDV(TM)Z8 Surgical Laser is intended for use in the creation of capsulotomy, phacofragmentation and the creation of single plane, multi-plane, arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.
Status
Terminated
Initiated
2018-01-21
Posted
not on file
Terminated
2018-11-02
Reason
This Correction is being initiated based on a customer complaint received on November 16, 2017 which indicated a complication with a cataract procedure. An investigation completed by the manufacturer determined that although the complication was due to user error improvements could and should be made on our part to improve the usability and performance of our device.
Root cause
Device Design
Action
The firm, Ziemer Ophthalmology, initiated their field correction by telephone on 01/21/2018 and followed with a letter on 01/28/2018. The letter described the product, problem and actions taken. The letter stated: "...we will be releasing new software which contains an added safety feature which will prevent unintended shift of the planned ARC Incisions and/or alert you in case of unforeseen eye movement between configuration and execution of the ARC incisions. We will be contacting your practice the week of March 12th to set up a time to install this new software and provide in-depth training. Prior to the installation of the new software, you may continue to use the Z8 for cataract surgery; however, we would advise that you not perform both arcuate and clear corneal incisions on the same patient until the new software is installed. If you have any questions, do not hesitate to contact Director of Operations at 618-462-9301 for any further information you require.
Quantity
125 devices
Distribution
Worldwide Distribution: US (nationwide) including states of: AL, CA, FL, GA, IL, IN, MN, OR, TN, TX, VA, and WA; and countries of: Argentina, Austria, Belgium, Brazil, Canada, France, Germany, India, Iraq, Israel, Italy, Japan, Lebanon, Malaysia, Mexico, Namibia, Pakistan, Peru, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Spain, South Africa, Switzerland, Taiwan, and United Arab Emirates.
Lot, serial or model codes
Serial Numbers:  FLM4455 FLM2313 FLM4499 FLM3230 FLM2855 FLM3097 FLM4493 FLM3178 FLM4441 FLM1592 FLM2595 FLM1225 FLM4542 FLM3771 FLM1464 FLM3999 FLM2408 FLM1995 FLM4409 FLM2185 FLM2679 FLM4987 FLM3815 FLM3904 FLM1958 FLM1179 FLM2869 FLM2584 FLM4860 FLM3766 FLM2318 FLM3406 FLM4626 FLM2324 FLM1326 FLM2053 FLM2096 FLM4222 FLM4450 FLM2266 FLM4588 FLM4175 FLM4764 FLM4902 FLM4227 FLM2457 FLM1767 FLM4132 FLM4896 FLM3007 FLM1274 FLM1009 FLM1320 FLM2537 FLM3278 FLM2508 FLM3905 FLM4043 FLM4133 FLM4314 FLM1819 FLM2186 FLM3135 FLM2627 FLM4271 FLM1092 FLM4759 FLM1724 FLM2589 FLM4770 FLM2038 FLM1188 FLM1640 FLM1915 FLM4317 FLM1814 FLM2812 FLM1003 FLM4859 FLM1135 FLM4807 FLM1502 FLM3183 FLM1182 FLM4954 FLM4000 FLM4531 FLM2551 FLM1554 FLM3539 FLM1046 FLM2765 FLM4897 FLM2913 FLM1093 FLM4632 FLM1686 FLM2912 FLM1782 FLM1545 FLM3359 FLM4908 FLM2085 FLM1407 FLM1863 FLM4308 FLM3677 FLM2636 FLM1008 FLM4361 FLM4669 FLM4498 FLM3729 FLM1773 FLM1364 FLM1591 FLM1824 FLM2229 FLM2450 FLM2722 FLM3676 FLM3861 FLM4593 FLM3040 FLM2001
510(k) numbers
["K150323"]
PMA numbers
not on file
Product code
OOE
Firm FEI
3005643720
Firm city
Port
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register