FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Recall number
- Z-1272-2014
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Status
- Terminated
- Initiated
- 2014-02-07
- Posted
- 2014-03-22
- Terminated
- 2014-12-30
- Reason
- Certain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop out of the gantry system.
- Root cause
- Device Design
- Action
- Siemens sent a recall notification letter, dated February 7, 2014, to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions call +49 9191 18 0.
- Quantity
- 37
- Distribution
- USA Nationwide Distribution in the states of WI, KY, OH, NJ, NC, FL, PA, MA, IA, TX, NY, IL, MI, AZ, CA, and LA.
- Lot, serial or model codes
- Model nos. 7412807, 7727717, 7728392 with serial numbers: 35020, 34999,35098 35113 35042 35061 35064 35073 35014 35069 35053 35055 35056 35054 35008 35057 35044 35018 35097 35108 35046 35048 35050 35045 35004 35019 46004 46005 40000 40009 40011 40006 40007 40010 44016 44005 44011
- 510(k) numbers
- ["K073290"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28