Baxter Healthcare Corp.: A) Product Code 2N8220: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.7", Vol. 1.10 mL, Luer Activated Valve for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; B) Product Code 2N8221: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.1", Vol. 0.85 mL, Injection Site Male Luer Lock Adapter with Retractable Collar, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; C) Product Code 2N8222: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 8.2", Vol. 0.57 mL, Luer Activated Valve for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; D) Product Code 2N8223: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.7", Vol. 0.40 mL, Injection Site Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; E) Product Code 6N8220: V-LINK Luer Activated Device with VITALSHIELD Protective Coating Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.6", Vol. 1.0 mL, Catheter Extension Set, Luer Activated Device with Silver for IV Access, Male Luer Lock Adapter with Retractable Collar, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; F) Product Code 6N8222: V-LINK Luer Activated Device with VITALSHIELD Protective Coating Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 8.2", Vol. 0.5 mL, Catheter Extension Set, Luer Activated Device with Silver for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips Product Usage: The IV extension tubing kits are intended to provide users with a ready to use set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The IV extension tubing sets are indicated for use with a vascular access device for administration and withdrawal of fluids.

Recall number
Z-1270-2013
Recalling firm
Baxter Healthcare Corp.
Product
A) Product Code 2N8220: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.7", Vol. 1.10 mL, Luer Activated Valve for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; B) Product Code 2N8221: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.1", Vol. 0.85 mL, Injection Site Male Luer Lock Adapter with Retractable Collar, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; C) Product Code 2N8222: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 8.2", Vol. 0.57 mL, Luer Activated Valve for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; D) Product Code 2N8223: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.7", Vol. 0.40 mL, Injection Site Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; E) Product Code 6N8220: V-LINK Luer Activated Device with VITALSHIELD Protective Coating Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.6", Vol. 1.0 mL, Catheter Extension Set, Luer Activated Device with Silver for IV Access, Male Luer Lock Adapter with Retractable Collar, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; F) Product Code 6N8222: V-LINK Luer Activated Device with VITALSHIELD Protective Coating Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 8.2", Vol. 0.5 mL, Catheter Extension Set, Luer Activated Device with Silver for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips Product Usage: The IV extension tubing kits are intended to provide users with a ready to use set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The IV extension tubing sets are indicated for use with a vascular access device for administration and withdrawal of fluids.
Status
Terminated
Initiated
2013-03-28
Posted
2013-05-07
Terminated
2017-07-31
Reason
Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
Root cause
Packaging process control
Action
Baxter sent an "Urgent Product Notification" dated March 28, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taked. Customers are advised to locate and remove all of the affected products from their inventory. If the products were further distributed, consigneees are requested to notify their customers of the recall. Products can be returned for credit by contacting Baxter. An acknowledgement form is required to be compelted by customers.
Quantity
A) Code 2N8220: 5,950; B) Code 2N8221: 6,150; C) Code 2N8222: 1,850; D) Code 2N8223: 7,900; E) Code 6N8220: 52,200; F) Code 6N8222: 1,200
Distribution
Worldwide Distribution - USA Nationwide and the countries of Puerto Rico, Brazil, Canada and Columbia.
Lot, serial or model codes
A) Product Code 2N8220: Lots R12J05070;   B) Product Code 2N8221: Lots R12J20046, R12J22042;   C) Product Code 2N8222: Lots R12L07031;   D) Product Code 2N8223: Lots R12L06074, R12K02083;   E) Product Code 6N8220: Lots R12I01054, R12H25048, R12H27192, R12J04057, R12J22109, R12L07049, R12L08088;   F) Product Code 6N8222: Lots R12J15046, R12J23115
510(k) numbers
["K103551"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register