Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: MOSAIQ System; customers with Siemens branded linear accelerators with a COHERENCE RT Therapist workspace version 2.1a and the 3rd party MOSAIQ Oncology Information System. MOSAIQ Oncology information system is owned and marketed by IMPAC Medical System, Inc. MOSAIQ is distributed by Siemens Medical Solutions USA, Inc, Radiation Oncology under a distributor contract with IMPAC Medical. Distributed by Siemens Healthcare, Concord, Ca. Product Usage: Medical charged particle radiation therapy system. The intended use of the SIEMENS branded PRIMUS, ONCOR and ARTISTE Solution family of linear accelerators systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. P

Recall number
Z-1270-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
MOSAIQ System; customers with Siemens branded linear accelerators with a COHERENCE RT Therapist workspace version 2.1a and the 3rd party MOSAIQ Oncology Information System. MOSAIQ Oncology information system is owned and marketed by IMPAC Medical System, Inc. MOSAIQ is distributed by Siemens Medical Solutions USA, Inc, Radiation Oncology under a distributor contract with IMPAC Medical. Distributed by Siemens Healthcare, Concord, Ca. Product Usage: Medical charged particle radiation therapy system. The intended use of the SIEMENS branded PRIMUS, ONCOR and ARTISTE Solution family of linear accelerators systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. P
Status
Terminated
Initiated
2012-02-14
Posted
2012-03-21
Terminated
2012-09-21
Reason
Images used in treatment planning may belong to another patient. Risk of treatment to wrong area. Software update to upgrade Siemens installed base to the latest release of MOSAIQ version 2.00X3 SP5, for RT Therapist version 2.1A
Root cause
Device Design
Action
Siemens sent a Customer Information letters to all affected customers. The Customer information letters inform customers of the release of MOSAIQ 2.00X3 SP5 which specifically addresses a safety issue that is only found when MOSAIQ is used with COHERENCE Therapist 2.1 MQ (also called RTT2.1A). This letter is to be kept with Customer information documentation. Update Instruction TH004/12S are being issued with the update. The Software update is delivered and installed by qualified Field Service Engineers. For questions, contact your responsible SIEMENS Service Engineer or Application Specialist.
Quantity
44 affected Siemens Customers
Distribution
Worldwide Distribution - USA (nationwide) including the states of KY, PA, TN, WI, and P.R. and the countries of Australia, Belgium, Bosnia Herzeg. China, Colombia, Croatia, Czech Republic, Germany, India, Ireland, Japan, Lebanon, Malaysia, Mexico, New Zealand, Philippines, Poland, Russian Federation, South Africa, Saudi Arabia, Thailand, Turkey, and United Kingdom.
Lot, serial or model codes
MOSIAQ System model number 10568605.
510(k) numbers
["K090683","K981313"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register