Baxter Healthcare Corp.: A) Product Code 2N3371: Non-DEHP Y-Type Catheter Extension Set, 5.7", Vol. 0.8 mL, 2 Injection Sites, Male Luer Lock Adapter, Single use only Product Usage: This device is indicated for use in blood sampling and the administration of solutions. The InterLink Injection Site feature is intended to eliminate accidental needle sticks when used in conjunction with an InterLink Cannula as part of an I.V. needless access system.
- Recall number
- Z-1269-2013
- Recalling firm
- Baxter Healthcare Corp.
- Product
- A) Product Code 2N3371: Non-DEHP Y-Type Catheter Extension Set, 5.7", Vol. 0.8 mL, 2 Injection Sites, Male Luer Lock Adapter, Single use only Product Usage: This device is indicated for use in blood sampling and the administration of solutions. The InterLink Injection Site feature is intended to eliminate accidental needle sticks when used in conjunction with an InterLink Cannula as part of an I.V. needless access system.
- Status
- Terminated
- Initiated
- 2013-03-28
- Posted
- 2013-05-07
- Terminated
- 2017-07-31
- Reason
- Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
- Root cause
- Packaging process control
- Action
- Baxter sent an "Urgent Product Notification" dated March 28, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taked. Customers are advised to locate and remove all of the affected products from their inventory. If the products were further distributed, consigneees are requested to notify their customers of the recall. Products can be returned for credit by contacting Baxter. An acknowledgement form is required to be compelted by customers.
- Quantity
- A) Product Code 2N3371: 228,200 units
- Distribution
- Worldwide Distribution - USA Nationwide and the countries of Puerto Rico, Brazil, Canada and Columbia.
- Lot, serial or model codes
- A) Product Code 2N3371: Lots R12H23076, R12H30097, R12I0707, R12I08059, R12I25046, R12I2607, R12J11052, R12J12068, R12J13058R12K17057, R12L10084, R13A12057
- 510(k) numbers
- ["K925126"]
- PMA numbers
- not on file
- Product code
- FPA
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28