FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc: Siemens ADVIA Centaur Vitamin D Diluent 2-pack, Reference 10494100 in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium heparin) using the ADVIA Centaur and ADVIA Centaur XP systems The ADVIA Centaur Vitamin D Total (Vit D) assay is for in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium heparin) using the ADVIA Centaur and ADVIA Centaur XP systems. The ADVIA Centaur Vit Assay is intended as an aid in the determination of vitamin D insufficiency. The ADVIA Centaur Vitamin D Diluent 2-pack is used on board the ADVIA Centaur systems to perform automatic and system-performed dilutions fro Vitamin D assay.

Recall number
Z-1268-2012
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
Siemens ADVIA Centaur Vitamin D Diluent 2-pack, Reference 10494100 in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium heparin) using the ADVIA Centaur and ADVIA Centaur XP systems The ADVIA Centaur Vitamin D Total (Vit D) assay is for in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium heparin) using the ADVIA Centaur and ADVIA Centaur XP systems. The ADVIA Centaur Vit Assay is intended as an aid in the determination of vitamin D insufficiency. The ADVIA Centaur Vitamin D Diluent 2-pack is used on board the ADVIA Centaur systems to perform automatic and system-performed dilutions fro Vitamin D assay.
Status
Terminated
Initiated
2012-02-16
Posted
2012-03-21
Terminated
2013-10-17
Reason
Automatic and system-performed dilution calculation factor for Vitamin D dilutions is incorrect and causes diluted patient samples to under recover by approximately 50%.
Root cause
Nonconforming Material/Component
Action
The firm, Siemens Healthcare Diagnostics, sent an "Urgent Device Recall Notice" on February 16, 2012 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to discontinue using the automatic and system-performed dilution features for Vitamin D; to discontinue the use of ADVIA Centaur VitD Diluent, 2-pack (REF 10494100); to manually dilute samples with values greater than the assay range of 150 ng/mL (375 nmoI/L), using the ADVIA Centaur VitD Diluent, 1 bottle (REF 10632114) as stated in the Vitamin D assay Instructions For Use (IFU); to review the contents of the recall notice with their Laboratory Director and retest the previous sample results generated using system- performed dilutions, and complete and return the confirmation fax-back form via fax to TECHNICAL SOLUTIONS CENTER at (302) 631-7597. If you have any questions or need additional information, please contact your local Technical Support Provider or Distributor or call (508) 668-5000.
Quantity
442 units
Distribution
Worldwide distribution: USA (nationwide) and countries of: Columbia, China, Denmark, Egypt, and Hong Kong.
Lot, serial or model codes
Lot # 80230 Expiry Date 04 May 2012
510(k) numbers
["K110586"]
PMA numbers
not on file
Product code
MRG
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register