Skip to the content

FDA Medical Device Recalls (openFDA)

Medica Corporation: EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.

Recall number
Z-1267-2018
Recalling firm
Medica Corporation
Product
EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.
Status
Terminated
Initiated
2017-04-11
Posted
not on file
Terminated
2019-08-20
Reason
Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
Root cause
Process design
Action
Consignees were notified of the recall via email bulletin dated 4/11/17. The bulletin informs customers of the issues and explains the actions to be taken.
Quantity
not on file
Distribution
USA: MI FL CA TX India, Russia, Puerto Rico, N. Mariana Island/Palau.
Lot, serial or model codes
Medica s Lot # 16292
510(k) numbers
["K120497"]
PMA numbers
not on file
Product code
JIT
Firm FEI
1220972
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register