FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Natural-Knee II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA TIB RT SZ0/16M ALL POLY ULTRA TIB LT SZ1 ALL POLY ULTRA TIB RT SZ1 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB RT SZ2 ALL POLY ULTRA TIB LT SZ3 ALL POLY ULTRA TIB RT SZ3 ALL POLY ULTRA TIB RT SZ5

Recall number
Z-1262-2014
Recalling firm
Zimmer, Inc.
Product
Natural-Knee II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA TIB RT SZ0/16M ALL POLY ULTRA TIB LT SZ1 ALL POLY ULTRA TIB RT SZ1 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB RT SZ2 ALL POLY ULTRA TIB LT SZ3 ALL POLY ULTRA TIB RT SZ3 ALL POLY ULTRA TIB RT SZ5
Status
Terminated
Initiated
2014-02-25
Posted
2014-03-20
Terminated
2015-03-24
Reason
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Root cause
Storage
Action
Beginning 2/25/2014 all distributors were notified via electronic mail. Hospital risk managers and distributors, surgeons with affected inventory will also be notified via courier or direct mail. The Urgent Lot Specific Recall Notice identified the issue and responsibilities. Distributors were asked to locate and remove the affected product in their territory, as well as identifying hospitals and surgeons who have previously implanted the affected product. Distributors were asked to return on-hand affected products to Zimmer and ensure all of their products are accounted for using the form provided in the letter and email it back to corporatequality.postmarket@zimmer.com. Hospital risk managers and surgeons were asked to locate affected product and return it to their Zimmer sales representative. Questions or concerns should be addressed to customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST Communications outside of the United States will occur approximately two weeks after the United States communications. 100 % effective checks will be done.
Quantity
3638 units
Distribution
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Lot, serial or model codes
Item Number 621800809 621801809 621801819 621801913 621801916 621810816 621810909 621820809 621820811 621820911 621830811 621830909 621850916
510(k) numbers
K936159
PMA numbers
not on file
Product code
JWH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source