Skip to the content

FDA Medical Device Recalls (openFDA)

Radiometer America Inc: Radiance Data Management System Software Version 2.42 (Stand-alone software package) for the Radiance STAT Analyzer Management System, in vitro diagnostic.

Recall number
Z-1262-06
Recalling firm
Radiometer America Inc
Product
Radiance Data Management System Software Version 2.42 (Stand-alone software package) for the Radiance STAT Analyzer Management System, in vitro diagnostic.
Status
Terminated
Initiated
2005-03-02
Posted
2006-07-27
Terminated
2012-02-24
Reason
Software anomaly. Data generated by blood gas equipment did not correlate with data shown in the Radiance Data Management System.
Root cause
Other
Action
The firm initiated this recall with letters to their customers on 3/2/2005. After this notification, the firm''s Field Service Reps visited each affected customer and updated their software to version 2.42. According to the firm, these actions took place from 3/12/2005 thru 4/15/2005. The Field Reps. of Radiometer America reported to Radiometer Medical ApS in Denmark that the necessary corrections had been competed as of: 4/15/2005.
Quantity
306 units
Distribution
Nationwide.
Lot, serial or model codes
Model 914-317 Radiance Basic Kit. Version 2.42 operating software
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1523456
Firm city
Westlake
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register