Skip to the content

FDA Medical Device Recalls (openFDA)

Radiometer America Inc: Dri-Tek BG-OX Cartridge, Model #D7120, Cat. #946-003 Cartridges used in the NPT7 Analyzer, an in vitro diagnostic device.

Recall number
Z-1261-06
Recalling firm
Radiometer America Inc
Product
Dri-Tek BG-OX Cartridge, Model #D7120, Cat. #946-003 Cartridges used in the NPT7 Analyzer, an in vitro diagnostic device.
Status
Terminated
Initiated
2006-05-01
Posted
2006-07-27
Terminated
2012-02-24
Reason
Air bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.
Root cause
Other
Action
The recalling firm initiated this recall with a letter to send to their customers, via Registered Mail, on 5/1/2006. The letter requests that the customers identify and discard the potentially defective cartridges from Lot #R0299. The letter is accompanied by a recall response page to be completed by the customer and faxed or mailed back to the Radiometer so that the customer can receive an equal number of replacement cartridges free-of-charge.
Quantity
185
Distribution
Nationwide.
Lot, serial or model codes
D7120 946-003 Cartridges, Lot #R0299, Expiration Date: 02/2008.
510(k) numbers
["K982928"]
PMA numbers
not on file
Product code
CHL
Firm FEI
1523456
Firm city
Westlake
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register