Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK
Status
Open, Classified
Initiated
2023-02-28
Posted
2023-03-16
Terminated
not on file
Reason
The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.
Root cause
Mixed-up of materials/components
Action
On 2/28/23, "URGENT MEDICAL DEVICE RECALL" notices were mailed and emailed to customers and distributors who were told to take the following actions: 1) Discontinue use of and Dispose of all affected product remaining in your possession in accordance with your local facility's process. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. 3) If you purchased this product from a distributor, contact your distributor for further instructions and credit resolution. 4) Complete and return the attached Customer Response Form and email to: BDRC3@bd.com The recalling firm will issue replacements, as available, or credit. In addition, distributors were asked to do the following: Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification. Customers/distributors who have questions can contact the North American Regional Complaint Center at 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted, M-F 8am - 5pm CT
Quantity
38,635
Distribution
Worldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO, PA, TX, NC, CA, ME, MO, NV, MI, IL, NY, VA, AZ, NH, LA, WI, KY, MA, WA, CT, SD, OK, DC, IA, AR, GA, NE, NM, IN, NJ, MN, PR, AL, AK, SC, KS, OR, ID, VT, MT, WV, ND, TN, MS, HI, WY, RI and OUS countries of: Canada, Korea, Thailand and Taiwan. OUS: AR, AU, BE, CA, CN, CO, HK, IN, JP, KR, TH, TW
Lot, serial or model codes
REF/UDI-DI/Lot(Expiration Date): 1610MSK/00801741142567/1445630(11/28/2024), 1449980(12/28/2024), 1451134(1/28/2025); 1816MSK/00801741097065/1447510(12/28/2024), 1447511(12/28/2024), 1447512(12/28/2024), 1447513(12/28/2024), 1448118(12/28/2024), 1448119(12/28/2024), 1448120(12/28/2024), 1448121(12/28/2024), 1448743(12/28/2024), 1450413(12/28/2024), 1450414(12/28/2024), 1451075(1/28/2025), 1451076(1/28/2025), 1451621(1/28/2025), 1451622(1/28/2025), 1451625(1/28/2025), 1451626(1/28/2025), 1453516(1/28/2025), 1453517(1/28/2025), 1454295(1/28/2025), 1456850(2/28/2025), 1456856(2/28/2025), 1457030(2/28/2025)
510(k) numbers
["K171953"]
PMA numbers
not on file
Product code
KNW
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register