Skip to the content

FDA Medical Device Recalls (openFDA)

Ethicon, Inc.: Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device Product Code STRAP25. Intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures, such as hernia repairs.

Recall number
Z-1260-2011
Recalling firm
Ethicon, Inc.
Product
Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device Product Code STRAP25. Intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures, such as hernia repairs.
Status
Terminated
Initiated
2010-12-29
Posted
2011-02-18
Terminated
2012-11-07
Reason
The Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device may have the potential for the sterile barrier to be compromised due to a defect in the packaging.
Root cause
Pending
Action
Ethicon sent Urgent Voluntary Product Recall letters dated, December 28, 2010, to their customers on December 29, 2010 via UPS. The letter identified the affected product and described the reason for recall. Customers are to discontinue use and sampling immediately. They are to return the product using the enclosed prepaid shipping label along with the enclosed business reply card regardless of whether they have any affected product or not. Customers should contact 1-877-551-7147 or speak to their manager if they have questions regarding this recall.
Quantity
4,110
Distribution
Worldwide Distribution -- USA and Canada.
Lot, serial or model codes
All Lots. Batch Numbers: CJM838,CJZ014, CJZ065, CJZ228, CJZ456, CJZ542, CJZ625, CJ6116, CJ6293, CJ6402, CJ6572, CJ6852, CJK043, CJK216, CJK330, CJK536, CJK659, CKM791, CKZ068, CKZ437, CKZ652, and CK6002.
510(k) numbers
["K093845"]
PMA numbers
not on file
Product code
GDW
Firm FEI
2210968
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register