Beckman Coulter Inc: Cytomics FC 500 Flow Cytometry System Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500 (100V)), 6605628 (TN; 5 CLR, FC500 (120V)) , 6605629 (TN; 5 CLR, FC500 (220V)), 6605630 (TN; 5 CLR, FC500 (240V)) with CXP Software Versions 2.0 and 2.1

Recall number
Z-1259-06
Recalling firm
Beckman Coulter Inc
Product
Cytomics FC 500 Flow Cytometry System Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500 (100V)), 6605628 (TN; 5 CLR, FC500 (120V)) , 6605629 (TN; 5 CLR, FC500 (220V)), 6605630 (TN; 5 CLR, FC500 (240V)) with CXP Software Versions 2.0 and 2.1
Status
Terminated
Initiated
2006-05-22
Posted
2006-07-26
Terminated
2012-04-26
Reason
Beckman Coulter has determined that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results.
Root cause
Other
Action
A Product Corrective Action letter was sent via US mail the week of May 22, 2006 to all FC500 customers. They were informed that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results. The letter includes the sequence in which the issue occurrs: 1) During the acquistion of a tube, the Pause and Rotate Button is selected. 2) A system error is generated during the Pause and Rotate routine (The error message table may be viewed by going to the CXP Help Menu, select the CXP Help Menu Item, then to Section 13.5). 3) An error message appears in the Cytometer Status, an audible alarm sounds, the run aborts, the carousel ejects, and the system automatically transitions from Verification to Run Initialization and then to Awaiting Sample 4) The Worklist is continued or a new Worklist is loaded and started. 5) After the Worklist is continued, or loaded and started, during the acquistion of a tube, the Pause and Rotate Button is selected. 6) The carousel Pauses but does not Rotate. 7) When the Start/Continue or Restart button is selected, the carousel may move to the wrong tube and aspirate sample resulting in a sample misidentification. If no system errors are generated during a Pause and Rotate routine, the system will function correctly. The letter also provides steps for immediate user preventive action if a system error is generated during the Pause and Rotate Routine to remedy the issue: 1) Exit CXP Software. 2) Power off the Cytometer using the FC OFF utility 3) Turn the Cytometer back on. 4) Recovery should be complete and the laboratory may resume normal usage of the system. Customers are asked to share the information with their laboratory staff and to retain this notification as part of their laboratory Quality System documentation and to complete and return the enclosed Fax Response Form for the firm''s records.
Quantity
424
Distribution
Nationwide and Canada
Lot, serial or model codes
Software Versions 2.0 and 2.1
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXG
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register