Skip to the content

FDA Medical Device Recalls (openFDA)

Tandem Diabetes Care Inc: t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology

Recall number
Z-1258-2022
Recalling firm
Tandem Diabetes Care Inc
Product
t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology
Status
Open, Classified
Initiated
2020-09-10
Posted
2022-06-10
Terminated
not on file
Reason
Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will continue for 30 minutes and then the pump will power off and insulin delivery will stop. If the battery reaches 1%, then insulin delivery will stop, which could lead to hyperglycemia.
Root cause
Software design
Action
On 09/10/2020, correction notices were sent to customers who were asked to acknowledge Low Power Alerts and Low Power Alarms by tapping "OK" and then charging the pump. Further, take the following precautions to ensure that your pump is properly charged, and the pump's battery life does not fall below 25% remaining power: - Periodically check the battery level indicator of your pump - Charge the pump for a short period of time every day (10 to 15 minutes) - Avoid full battery discharges A software update is planned by the end of 2020 that resolves the potential fluctuating battery life display. We will reach out to you to complete the free remote update of your insulin pump when the software update becomes available. Complete a online form to confirm that you have received this important user communication. Customers with questions were encouraged to contact diabetes@amsl.com.au In November 2020 software version 6.4.1 was released for all affected insulin pumps except Dexcom G5 Mobile CGM system pumps. On 12/06/2020, Dexcom G5 Mobile CGM system customers were informed that a new software update was available to address the battery issues but that the update requires using the Dexcom G6 CGM system. On 5/24/22, correction notices were sent to customers in Australia to update to the latest software. t:slim X2 insulin pump with Basal-IQ technology version 6.6 and Control-IQ technology version 7.6 are available with the latest software versions. Software update contact: AMSL: diabetes@amsl.com.au or 1300 851 056.
Quantity
5,206
Distribution
International distribution in the country of Australia.
Lot, serial or model codes
Dexcom G5 - All Serial Numbers; Basal-IQ Technology - Software Versions 6.3.01 and Earlier
510(k) numbers
not on file
PMA numbers
not on file
Product code
PQF
Firm FEI
3007981285
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register