Toshiba American Medical Systems Inc: Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner

Recall number
Z-1257-2011
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner
Status
Terminated
Initiated
2010-09-07
Posted
2011-02-15
Terminated
2012-04-20
Reason
Incorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is approximately 10% lower than what the displayed value should be when the scanner executes the exam Plan.
Root cause
Software design
Action
Toshiba America Medical Systems (TAMS) issued notification to consignees 9/7/2010, via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.
Quantity
44 units distributed between 09/2007 - 07/2010
Distribution
Nationwide distribution.
Lot, serial or model codes
Affected Serial No.: 2DA0872038, 2DA0962116, 2DA0862034, 2DA09Z2161, 2DA07X2008, 2DA1012165, 2DA08X2072, 2DA09Y2151, 2DA0842023, 2DA07X2005, 2DA1012168, 2DA08X2067, 2DA0882053, 2DA0862036, 2DA08Z2092, 2DA0882051, 2DA09Z2162, 2DA1072211, 2DA09X2143, 2DA08Y2077,2DA09X2139, 2DA08X2070, 2DA0982133 2133, 2DA0842026, 2DA0882052, 2DA0972123, 2DA08Y2080, 2DA0882048, 2DA08X2063, 2DA09Y2148, 2DA1012170, 2DA08X2071, 2DA1012169, 2DA0952115, 2DA0982131, 2DA07Y2010, 2DA08X2061, 2DA1012167, 2DA0882049, 1DA0962006, 1DA09Y2014, 1DA0992011, 1DA0972007, 1DA0972008, 2DA1012166, 2DA09Z2158, 2DA1072212, 2DA1012174, 2DA0792004, 2DA1062208, 2DA0792002, 1DA1062015, 1DB1072012
510(k) numbers
["K083282"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register